Command Descriptions
ABORt (No Query Form)
The ABORt command terminates a measurement. The trigger subsystem state
is set to idle-state. The command does not invalidate already
fi
nished results
when breaking an array measurement. This means that you can fetch a partial
result after an abort.
Aborts all previous measurements if *WAI is not used.
Group
Measurement
Syntax
ABORt
ACQuisition:APERture
Sets the gate time for one measurement.
Group
Sense
Syntax
ACQuisition:APERture {<Decimal value > | MIN | MAX }
ACQuisition:APERture?
Arguments
<DECIMAL VALUE>
is 20 ns to 1000s.
MIN
sets 20 ns and
MAX
sets 1000 s.
Returns
<Decimal value >
200 ms after SYST:PRES
10 ms after *RST
ACQuisition:HOFF
Sets the Hold Off function On or Off.
Group
Sense
Syntax
ACQuisition:HOFF <boolean>
ACQuisition:HOFF?
FCA3000, FCA3100, MCA3000 Series Programmer Manual
2-29
Содержание FCA3000 Series
Страница 2: ......
Страница 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 135: ...Status and Events ...
Страница 136: ......
Страница 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 157: ...Programming Examples ...
Страница 158: ......
Страница 179: ...Appendices ...
Страница 180: ......
Страница 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Страница 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...