Table 3-8: Query errors
Error Number
Error Description
Explanation and examples
–400
Query error
This code indicates that a Query Error as de
fi
ned in IEEE-488.2, 11.5.1.1.7
and 6.3 has occurred.
Query INTERRUPTED
Indicates that a condition causing an INTERRUPTED Query error occurred
(see IEEE-488.2, 6.3.2.3); for example, a query was followed by DAB or
GET before a response was completely sent.
Query INTERRUPTED; in
send state
Query INTERRUPTED; in
query state
–410
Query INTERRUPTED; in
response state
The additional information indicates the IEEE-488.2 message exchange
state where the error occurred.
Query UNTERMINATED
Indicates that a condition causing an UNTERMINATED Query error occurred
(see IEEE-488.2, 6.3.2.2); for example, the instrument was addressed to
talk and an incomplete program message was received.
Query UNTERMINATED; in
idle state
Query UNTERMINATED; in
read state
–420
Query UNTERMINATED; in
send state
The additional information indicates the IEEE-488.2 message exchange
state where the error occurred
–430
Query DEADLOCKED
Indicates that a condition causing an DEADLOCKED Query error occurred
(see IEEE-488.2, 6.3.1.7); for example, both input buffer and output buffer
are full and the instrument cannot continue.
–440
Query UNTERMINATED after
inde
fi
nite response
Indicates that a query was received in the same program message after an
query requesting an inde
fi
nite response was executed (see IEEE-488.2,
6.5.7.5.7.)
Table 3-9: Device-speci
fi
c errors
Error Number
Error Description
Explanation and examples
(1)100
Device operation gave
fl
oating-point under
fl
ow
A
fl
oating-point error occurred during a instrument operation.
(1)101
Device operation gave
fl
oating-point over
fl
ow
A
fl
oating-point error occurred during a instrument operation.
(1)102
Device operation gave ‘not a
number’
A
fl
oating-point error occurred during a instrument operation.
(1)110
Invalid measurement function
The instrument was requested to set a measurement function it could not
make.
(1)120
Save/recall memory protected
An attempt was made to write in a protected memory.
FCA3000, FCA3100, MCA3000 Series Programmer Manual
3-17
Содержание FCA3000 Series
Страница 2: ......
Страница 8: ...Preface iv FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 9: ...Getting Started ...
Страница 10: ......
Страница 14: ...Using the USB Interface 1 4 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 15: ...Syntax and Commands ...
Страница 16: ......
Страница 26: ...Command Syntax 2 10 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 135: ...Status and Events ...
Страница 136: ......
Страница 156: ...Status and Events 3 20 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 157: ...Programming Examples ...
Страница 158: ......
Страница 179: ...Appendices ...
Страница 180: ......
Страница 181: ...Appendix A Character Set FCA3000 FCA3100 MCA3000 Series Programmer Manual A 1 ...
Страница 182: ...Appendix A Character Set A 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...
Страница 188: ...Appendix D Reserved Words D 2 FCA3000 FCA3100 MCA3000 Series Programmer Manual ...