Maintenance
Power Supply Module
Follow these steps to remove the power supply module. You need to have
previously removed the rear case assembly rear chassis, and I/O board. Use a
TORX T15 screwdriver tip. (See Figure 5-3 on page 5-9.)
1.
Disconnect the power cable from J104 on the I/O board (if still attached)
and from J0403 on the Main board.
2.
Disconnect the cable from the Power Supply module to J0401 on the Main
board.
3.
Disconnect the power connector ground lead from the chassis.
4.
Remove the
fi
ve T15 screws that attach the Power Supply module to the
chassis. When reinstalling torque these screws to 8.0 in/lb.
5.
Lift the Power Supply module, including the power connector, off of the
chassis and place it on a static-free surface.
MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual
4–9
Содержание DPO3000 Series
Страница 2: ......
Страница 12: ...Service Safety Summary viii MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 14: ...Preface x MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 15: ...Operating Information ...
Страница 16: ......
Страница 18: ...Operating Information 1 2 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 19: ...Theory of Operation ...
Страница 20: ......
Страница 24: ...Theory of Operation 2 4 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 25: ...Adjustment Procedures ...
Страница 26: ......
Страница 37: ...Maintenance ...
Страница 38: ......
Страница 56: ...Maintenance 4 18 MSO3000 and DPO3000 Series Digital Phosphor Oscilloscopes Service Manual ...
Страница 57: ...Replaceable Parts List ...
Страница 58: ......
Страница 77: ...Index ...
Страница 78: ......