Cursor function
Measurements
Horizontal cursors measure vertical parameters (amplitude) in volts,
watts, or IRE. The cursor readouts are defined as:
V1 = Level @ Cursor 1 with respect to its source ground level.
V2 = Level @ Cursor 2 with respect to its source ground level.
ΔV = Level @ Cursor 2 - Level @ Cursor 1.
Level is cursor displacement from the source ground multiplied by
the source in volts/div. The two cursors may have different sources
and therefore can have different volts/div settings.
Vertical cursors measure horizontal parameters. Typically the
horizontal parameters are in time or frequency, but depend on the
units you select. The cursor readouts are defined as:
T1 = Time @ Cursor 1 with respect to the trigger point
T2 = Time @ Cursor 2 with respect to the trigger point
ΔT = Time @ Cursor 2-Time @ Cursor 1
Time is divisions of displacement of the cursor from its source
trigger point multiplied by the source in time/div.
Waveform cursors measure vertical parameters (typically volts) and
horizontal parameters (typically time or frequency) simultaneously.
Each cursor is, in effect, both a vertical and horizontal cursor. These
paired cursors cannot be moved off of the waveform. The two
cursors can have a different source.
Use screen cursors to indicate an arbitrary X and Y position in the
waveform display area represented by the intersection of a vertical
and horizontal cursor line. Screen cursors measure both vertical
parameters (typically volts) and horizontal parameters (typically time
or frequency).
What do you want to do next?
Go to a step-by-step procedure for taking automatic measurements.
Histograms
The instrument can display histograms constructed from the selected waveform data. You can display both vertical (voltage) and
horizontal (time) histograms, but only one at a time. Use histogram measurements to get statistical measurement data for a
section of a waveform along one axis.
The histogram source can be any channel, math, or reference waveform.
Turning on histograms starts histogram counting and data accumulation. A sample histogram display is shown below. Histogram
data is continuously accumulated until you explicitly turn off histograms. This allows you to continue collecting histogram data
even when you turn off the histogram display.
In addition to using limit controls to set histogram box boundaries, you can also use standard Windows drag-and-drop to resize
and reposition the histogram box.
Histograms are not available in Zoom modes.
Go to a step-by-step procedure for starting histogram counting.
Go to a step-by-step procedure for taking automatic measurements.
Oscilloscope reference
DPO70000SX, MSO/DPO70000DX, MSO/DPO70000C, DPO7000C, and MSO/DPO5000B Series
701
Содержание DPO2304SX
Страница 1: ...Tektronix Digital Phosphor Oscilloscopes Printable Help P077006219 077 0062 19...
Страница 2: ......
Страница 3: ...Tektronix Digital Phosphor Oscilloscopes Printable Help www tek com 077 0062 19...
Страница 32: ...Introduction xxviii DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 83: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 51...
Страница 89: ...About Tektronix oscilloscopes DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 57...
Страница 100: ...About Tektronix oscilloscopes 68 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 130: ...Cursor setups 98 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 198: ...MagniVu setup 166 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 212: ...Display setups 180 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 234: ...Horizontal and acquisition setups 202 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 276: ...Mask testing setups 244 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 300: ...Math setups 268 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 408: ...MyScope setups 376 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 510: ...Trigger setups 478 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 548: ...Vertical setups 516 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 558: ...Copying and printing 526 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 605: ...Utilities DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 573...
Страница 606: ...Utilities 574 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 687: ...Figure A Figure B Oscilloscope reference DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series 655...
Страница 898: ...Oscilloscope reference 866 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...
Страница 920: ...Index 888 DPO70000SX MSO DPO70000DX MSO DPO70000C DPO7000C and MSO DPO5000B Series...