Operating
Instructions
—DC
502
OPERATIONAL
CHECK
Display
Check
Press
the
RESET button to check the 7 character
segments
of
each digit; the
numerical display should be a
row
of eights. To check the decimal
point position and the
units
indicators, set
the MEASUREMENT INTERVAL
switch
as follows:
Switch
Position
.01
SEC I
.1
SEC
1
_
(
DIRECT
1 SEC /
.
I
INPUT
10 SEC I
MANUAL
/
MANUAL 1
1
M0
10
SEC
I
.
I
PRE-
1 SEC >
.1
SEC
SCALE
1
INPUT
.01
SEC 1
Numerical
Display
Units
.0000
MHz
.00000
MHz
.000
kHz’
.0000
kHz
000
000
.000
kHz
.00000
MHz
.0000
MHz
.000
MHz
In
the MANUAL position, no decimal point will be
displayed.
Press
the
START button
and check that the
GATE
indicator lights, then release the button (STOP) and
check
that the GATE light goes out. To check the
OVERFLOW
indicator, set the MEASUREMENT
INTERVAL switch
to 10 s
and apply
15- or 20-MHz to the
INPUT
connector. The length of time a display can be held
is
determined by
the DISPLAY TIME control, and will be
discussed
in
the next few paragraphs.
Then
turn
the MEASUREMENT INTERVAL switch to
the
position
that gives
the
desired reading. Generally, use the
shorter
measurement intervals for high-frequency, low-
resolution
measurements
and the longer intervals for
measurements
requiring
a high resolution.
NOTE
The
OVERFLOW
indicator can be lit for high-
resolution
measurements, allowing
the frequency to
be
indicated to 0.1
Hz.
Refer to
the Specifications at
the
end of this section
for resolution and accuracy at
each position
of the MEASUREMENT
INTERVAL
switch.
The
display is updated
at
a
rate determined
by the
DISPLAY TIME
control. Each time a
sample of the input
signal is
taken,
the
GATE light will flash and
the new
reading
will
be
displayed. To change the display time,
which
is continuously variable from
about 0.1 second to 10
seconds,
or to
hold a display indefinitely, turn the
DISPLAY
TIME
control.
÷10
Pre-Scale
Input.
The
DC 502 also provides
a pre
scaled,
AC-coupled input
to measure the average frequency
of
signals from
50
MHz to 550 MHz. This input has a
sensitivity
of 500
mV,
peak-to-peak, and a maximum input
voltage
limit of 10 V, peak-to-peak. Signals greater than
10
V may damage the diodes of the
input circuit. The
ATTEN
controls do
not apply
to this input.
Frequency Measurements
Direct
Input.
The
DC 502
provides
direct measurement
of
the average frequency of signals from about 10 Hz to
110
MHz. The input sensitivity is 300 mV peak-to-peak, so
select
the
proper attenuation
(X1, X5, X10, or X50) for the
given signal.
The input signa! must not exceed 500 volts.
Apply
a
50 MHz to 550
MHz signal
of
at least 500 mV
amplitude
to
the ÷10 PRE-SCALE INPUT. Set the
MEASUREMENT
INTERVAL switch to .01 SEC and
observe
the readout.
Leading
zeros
should
be
blanked.
Select a MEASUREMENT
INTERVAL which gives the best
accuracy
and resolution. As with the DIRECT INPUT,
shorter measurement
intervals
give higher frequency, lower
resolution
measurements; longer
intervals offer greater
resolution, especially
when overflow is employed.
Apply
a
signal to the INPUT connector. Set the
MEASUREMENT INTERVAL switch to the .01
SEC posi
tion
and the DISPLAY TIME
control fully CCW. Observe
the
numerical
readout
display. Adjust
the TRIGGER
LEVEL
control for a stable reading. The zeroes leading the
most
significant digit
in the display should be blanked.
Totalizing
DIRECT
INPUT.
The
DC 502
will count and display the
accumulated number
of signals (events) applied to the
DIRECT
INPUT connector up to
the
register capacity of
9,999,999
during the time interval between START/STOP
commands
from
the front-panel pushbutton. Input
signal
rate should
not exceed 110 MHz.
®ï
1-2
REV.
MAY 1974
Содержание DC 502
Страница 4: ...DC 502...
Страница 24: ...39 39 Relocated on back of board SN B070000 REV MAY 1974...
Страница 33: ...FIG 1 EXPLODED...