Tektronix AWG5200 Series Скачать руководство пользователя страница 16

d. Select the Calibration button, then the Calibration check box to select all self-calibrations, and change the Log options as

desired. All available tests and adjustments are selected now.

3. Click Start to begin the calibration. The Start button changes to Abort while the calibration is in process.
4. During calibration, you can click the Abort button to stop the calibration and revert to the previous calibration data. If you do so, no

calibration constants will be saved.

5. If you allow the calibration to complete, and there are no errors, the new calibration data is applied and saved. The pass/fail result is

shown in the right panel of the Calibration page, and contains the associated date, time, and temperature information.

6. Calibration data is automatically stored in non-volatile memory. If you don’t want to use the calibration data from the most recent

self-calibration, click the Restore factory cal button. This loads the original calibration data shipped with the instrument.

Diagnostics

This section contains information designed to help troubleshoot the AWG5200 series instruments to the module level. Component-level
repair is not supported.

Use the instrument diagnostics to help troubleshoot these instruments.

Note: The diagnostics are available during normal start-up of the AWG5200 Series application.

Back up data

Before running any diagnostics or calibrations on a unit, copy 

C:\ProgramData\Tektronix\AWG\AWG5200\Logs

 to a

another location.

Review this data with an XML editor or Excel spreadsheet to locate the errors. Then when you run diagnostics or calibrations, you can
compare current and previous instrument behavior.

Saving the persistence file

Before you start troubleshooting, use Microsoft Windows Explorer to back up the persistence file to a safe
service copy location. After service is complete, restore the persistence file. The persistence file location is

C:\ProgramData\Tektronix\AWG\AWG5200\persist.xml.

resultStatistics log file

The resultStatistics log file is a good starting point when diagnosing a reported problem. This file contains the instrument’s identification
data and includes which tests were run and results.

This is an .xml file and the best way to view the file is as follows:

1. Open a blank Excel spreadsheet.
2. Click on the Data tab.
3. Click Get Data and then select File > From XML.
4. Navigate to 

C:\ProgramData\Tektronix\AWG\AWG5200\resultStatistics.xml

 and import the data.

Diagnostics overview

The instrument runs some self-tests at start-up. These are the POST tests.

The POST tests check connectivity between the boards and also check that the power is within the required range, and that the clocks are
functional. You can also choose to run the POST tests at any time, by selecting POST Only in the Diagnostics window.

If there is an error, the instrument automatically goes into diagnostics.

The levels of diagnostics in the tree are:

Maintenance

AWG5200 Series Arbitrary Waveform Generators Service Manual 

16

Содержание AWG5200 Series

Страница 1: ...AWG5200 Series Arbitrary Waveform Generators Service Manual Register now Click the following link to protect your product www tek com register P 077174500 077 1745 00...

Страница 2: ...ued and pending Information in this publication supersedes that in all previously published material Specifications and price change privileges reserved TEKTRONIX and TEK are registered trademarks of...

Страница 3: ...g 12 Exterior inspection 12 Exterior cleaning 12 Lubrication 13 Remove and replace 13 Remove and replace procedures not requiring factory calibration 13 Software upgrades 15 Calibration 15 Diagnostics...

Страница 4: ...the country of use Use proper power cord Use only the power cord specified for this product and certified for the country of use Do not use the provided power cord for other products Use proper voltag...

Страница 5: ...damaged or operates incorrectly If in doubt about safety of the product turn it off and disconnect the power cord Clearly mark the product to prevent its further operation Before use inspect voltage...

Страница 6: ...depend on the probe and your application Refer to the Specifications section of the manual for more information Warning To prevent electrical shock do not exceed the maximum measurement or maximum fl...

Страница 7: ...rotective panels soldering or replacing components Verify safety after repair Always recheck ground continuity and mains dielectric strength after performing a repair Terms in the manual These terms m...

Страница 8: ...als all capitals etc in the manual as is used on the instrument front panel and menus Instruction steps are numbered unless there is only one step Bold text refers to specific interface elements that...

Страница 9: ...loads AWG5200 Series Rackmount Instructions GF RACK3U 071 3534 XX This document provides instructions to mount the AWG5200 Series Arbitrary Waveform Generators into a standard 19 inch equipment rack w...

Страница 10: ...receive clock and trigger timing and play out waveform data via an eight lane high speed serial interface JESD204B to the DAC The DAC creates the waveform The DAC output has four different paths DC Hi...

Страница 11: ...ge from electrostatic discharge Static voltages of 1 kV to 30 kV are common in unprotected environments CAUTION Static discharge can damage any semiconductor component in this instrument Observe the f...

Страница 12: ...s a guide An instrument that appears to have been dropped or otherwise abused should be checked thoroughly to verify correct operation and performance Immediately repair defects that could cause perso...

Страница 13: ...us injury or death Required equipment The following table lists equipment that you will need to remove and replace instrument modules Table 3 Required tools for removal and replacement of modules Name...

Страница 14: ...e holes in the handle with the posts on the instrument Attach the handle with two T25 screws and torque to 20 in lbs Side handle 1 Remove the four screws using T20 bit to remove the two handle top cap...

Страница 15: ...onnel If the front panel or rear panel is opened a full factory calibration must be performed by Tektronix Restore Factory calibration If you run a self cal and the results are bad you can restore the...

Страница 16: ...on a unit copy C ProgramData Tektronix AWG AWG5200 Logs to a another location Review this data with an XML editor or Excel spreadsheet to locate the errors Then when you run diagnostics or calibration...

Страница 17: ...Diagnostics Calibration you open a window where you can run the Self Calibration or Diagnostics The screen shows the last time that calibration ran and the internal temperature of the instrument when...

Страница 18: ...mensions six or more inches greater than the dimensions of the instrument Use a shipping carton that has a test strength of at least 50 pounds 23 kg b Surround the module with a protective anti static...

Страница 19: ...rd accessories for these products are listed in your user manual The user manual is available at www tek com manuals Changes to Tektronix products are sometimes made to accommodate improved components...

Страница 20: ...ction contains a list of the replaceable mechanical and or electrical components Use this list to identify and order replacement parts The following table describes each column in the parts list Parts...

Страница 21: ...er to Figure 1 on page 21 1 348 2037 XX 4 FOOT REAR CORNER SAFETY CONTROLLED 2 211 1481 XX 4 SCREW MACHINE 10 32X 500 PANHEAD T25 WITH BLUE NYLOK PATCH 3 211 1645 XX 2 SCREW MACHINE 10 32X 750 FLATHEA...

Страница 22: ...OK PATCH 9 348 2199 XX 4 CUSHION FOOT SANTOPRENE 4 BLACK 101 80 10 211 1645 XX 6 SCREW MACHINE 10 32X 750 FLATHEAD 82 DEG TORX 20 WITH THREAD LOCKING PATCH 11 367 0599 XX 2 HANDLE ASSY BASE AND GRIP S...

Отзывы: