2. Remove the rubber plug that is installed in the bottom foot you are replacing.
3. Remove the screw attaching the foot, and then remove the foot.
4. To replace the foot, place it in position and install the screw, using a T-20 tip, and torque to 10 in-lbs.
Handles
1. To remove the handles, place the instrument bottom on the work surface.
2. Remove the three screws that attach the handle to the instrument as shown, and remove the handle.
3. To replace the handles, position the handle on the instrument, lining up the holes in the handle with the posts on the instrument. Attach
the handle with two T25 screws and torque to 20 in-lbs.
Side handle
1. Remove the four screws using T20 bit to remove the two handle top caps. While installation, torque to 20 in*lb with T20 bit.
2. Remove the silicone handle from top of the spacers and remove the two spacers.
3. To replace, reverse the procedure.
Encoder knob
Note: The encoder knob is a push-button knob. You must leave at least 0.050 inches of clearance between the back face of the
knob and the front panel.
1. To remove the encoder knob, loosen the set screw. Do not remove the spacer and nut under the knob.
Maintenance
AWG5200 Series Arbitrary Waveform Generators Service Manual
14