Operating basics
6.
Select Burst as the Run
Mode for both the master
and the slave instrument.
To synchronize two
instruments, you must
stop running output
signals once before
triggering using the Burst
mode.
7.
Select the trigger source
of the slave unit.
Push the -more- bezel
button in the Burst mode
to display the second
page.
8.
Push the Source bezel
button to select External.
Con
fi
rm that Internal is
selected in the master
and External is selected
in the slave for Source.
9.
This screen shows signal
outputs by master-slave
operation. The top
waveform is the signal of
the master unit.
10.
When one meter long
cables are used, the
slave signal(s) delays
approximately 40 ns.
11.
To make the delay
between the master and
slave 0 (zero), set Delay
of the master unit.
12.
The screen shows
waveforms after setting
the delay to the master
unit. The top waveform
is master and the two
bottom waveforms are
slaves.
76