![Tektronix 80A00 Скачать руководство пользователя страница 71](http://html1.mh-extra.com/html/tektronix/80a00/80a00_service-manual_1077948071.webp)
Removal and installation
Figure 17: Sample of electrical module hardware removal
Replacing the chassis
CAUTION.
Transport the sampling module chassis with terminations on the inputs.
Transport the sampling module chassis in the original container and store the
sampling modules on a metal or conductive surface.
When handling the exposed chassis sampling module only, do not touch any
exposed component on the circuit board. Hold the exposed module by the edges of
the circuit board or the rear or front housings.
1.
Reinstall the chassis into the cover by reversing the removal procedure. (See
page 54,
WARNING.
Do not install or remove any sampling modules while the instrument
is powered on. Electrical shock may occur. Always power the instrument down
before attempting to remove or insert any sampling module to avoid potential
injury from shock.
2.
Install the electrical sampling module into one of the instrument small module
compartments.
DSA8300 Digital Serial Analyzer and Modules Service Manual
55
Содержание 80A00
Страница 2: ......
Страница 14: ...Compliance information x DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 28: ...Theory of operation 12 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 90: ...Replaceable parts list Figure 20 External parts 74 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 92: ...Replaceable parts list Figure 21 Drives 76 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 96: ...Replaceable parts list Figure 23 ATX assembly 80 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 108: ...Replaceable parts list End of service manual 92 DSA8300 Digital Serial Analyzer and Modules Service Manual ...