Removal and installation
1.
To access the assembly, remove these assemblies:
Front-Panel trim (See page 29.)
Top trim (See page 29.)
2.
To remove the Display assembly:
a.
Place the bottom of the instrument on the work surface, with the front
panel facing you.
b.
Remove the four T-15 Torx drive screws that attach the Display assembly
to the chassis.
c.
Hold the top and bottom edges of the Display assembly and pull forward
far enough to allow access to the ribbon cable connector.
d.
Disconnect the J2 and J3 ribbon cables from the Display assembly.
CAUTION.
Do not set the Display assembly on a work surface. Sliding the
instrument over the edge of the work surface could break off the Standby/On
switch assembly.
e.
Remove the Display assembly from the instrument.
f.
To reinstall the Display assembly:
Reverse the steps. Refer to the cable
connections table for information about where to connect cables. (See
page 72,
DSA8300 Digital Serial Analyzer and Modules Service Manual
35
Содержание 80A00
Страница 2: ......
Страница 14: ...Compliance information x DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 28: ...Theory of operation 12 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 90: ...Replaceable parts list Figure 20 External parts 74 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 92: ...Replaceable parts list Figure 21 Drives 76 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 96: ...Replaceable parts list Figure 23 ATX assembly 80 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 108: ...Replaceable parts list End of service manual 92 DSA8300 Digital Serial Analyzer and Modules Service Manual ...