![Tektronix 80A00 Скачать руководство пользователя страница 30](http://html1.mh-extra.com/html/tektronix/80a00/80a00_service-manual_1077948030.webp)
Adjustment procedures
Equipment hookup
1.
Connect the equipment as shown in the following
fi
gure:
2.
Set the National Instruments GPIB Interface command software to enable
GPIB communications between the PC controller and the DSA8300.
3.
Allow the test equipment and DSA8300 to warm up for at least 20 minutes
before starting the adjustment procedures.
Main instrument adjustments
Compensate the DSA8300
CAUTION.
To prevent incorrect calibration values, make sure that the instrument
and test equipment have warmed up for at least 20 minutes before performing
this procedure.
1.
Click
Utilities > Compensation
.
2.
Select
Mainframe
.
3.
Select
Compensate and Save
.
4.
Click
Execute
.
5.
Click
OK
in dialog box.
6.
Click
Close
when the compensation procedure is complete.
14
DSA8300 Digital Serial Analyzer and Modules Service Manual
Содержание 80A00
Страница 2: ......
Страница 14: ...Compliance information x DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 28: ...Theory of operation 12 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 90: ...Replaceable parts list Figure 20 External parts 74 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 92: ...Replaceable parts list Figure 21 Drives 76 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 96: ...Replaceable parts list Figure 23 ATX assembly 80 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Страница 108: ...Replaceable parts list End of service manual 92 DSA8300 Digital Serial Analyzer and Modules Service Manual ...