W A R N I N G
High voltage is present inside this instrument. To
avoid electric-shock hazard, operating personnel
must n o t remove the protective instrument
cover. Component replacement and internal ad
justments must be made b y qualified personnel
only.
In the ac power source mode, the 305 DMM Oscillo
scope operates from a single phase power source, which
has one o f its current carrying conductors at ground (earth)
potential. Operation from other power sources where both
current carrying conductors are live w ith respect to ground
(such as phase-to-phase on a multi-phase three wire system I
is n o t recommended because only the line conductor has
over-current (fuse) protection w ith in the instrument.
This instrument has a 3-wire power cord w ith a 3-
contact plug for connection to the power source and to
protective ground. The plug protective-ground contact con
nects (through the cord protective grounding conductor) to
the accessible metal parts o f the instrument. For electric-
shock protection, insert this plug in to a socket o u tle t that
has a securely grounded protective-ground contact.
L)o not defeat the grounding connection. A ny interrup
tion o f the grounding connection can create an electric-
shock hazard. Before making external connections to this
instrument, always ground the instrument firs t by connect
ing the power cord to a proper mating power outlet, that
is known to be properly grounded.
OPERATING VOLTAGE
This instrument may operate from an external dc source,
rechargeable batteries (supplied w ith the instrument), or a
115 or 230 vo lt ac nominal line voltage.
External DC Power Source Operation
The 305 can operate from an external dc power source
o f +9 to +32 volts. Set the Power Source Selector switch to
the EXT DC position (Figure 1). A pply external dc voltage
to the tw o banana jack inputs using the cable assembly
supplied w ith the instrument.
Internal Battery Power Source Operation
To operate the instrument from the internal battery
source, set the Power Source Selector switch to the B AT
TERY position (Figure 1). Battery voltage is indicated on
the DMM digital readout when all DMM FUNCTION push
buttons are in the o u t position.
2
305 Operators
REV A SEP 1979