![Tektronix 2215A Скачать руководство пользователя страница 70](http://html.mh-extra.com/html/tektronix/2215a/2215a_instruction-manual_1078831070.webp)
Adjustment Procedure— 2215A Service
m. Set the CH 2 VOLTS/DIV switch to 5 mV.
n. Repeat parts i through k for Channel 2.
o. Disconnect the test equipment from the instrument.
NOTE
Install the instrum ent cabinet fo r the remaining vertical
checks and allow a 20-minute warm-up period before
continuing with the Adjustm ent Procedure. See the
“Cabinet" remove and replace instructions located in
the “Maintenance" section o f the manual.
13. Check Bandwidth Limit Operation
a. Set:
VERTICAL MODE
CH 1
Vertical POSITION (both)
Midrange
BW LIMIT
On (button in)
VOLTS/DIV Variable (both)
CAL detent
Input Coupling (both)
DC
A SEC/DIV
20
ns
b. Connect the leveled sine-wave generator output via a
50-fi cable and a 50-fi termination to the CH 1 OR X input
connector.
c. Set the generator to produce a 50-kHz, 6-division
display.
d. Increase the generator output frequency until the dis
play amplitude decreases to 4.2 divisions.
e. CHECK— Generator output frequency is between
8.5 MHz and 11.5 MHz.
14. Check Bandwidth
a. Set:
BW LIMIT
Off (button out)
VOLTS/DIV (both)
2 mV
b. Set the generator to produce a 50-kHz, 6-division
display.
c. CHECK— Display amplitude is 4.2 divisions or greater
as the generator output frequency is increased up to the
value shown in Table 5-5 for the corresponding VOLTS/DIV
switch setting.
Table 5-5
Settings for Bandwidth Checks
VOLTS/DIV
Generator
Switch Setting
Output Frequency
2 mV
50 MHz
5 mV to 5 V
60 MHz
d. Repeat parts b and c for all indicated CH 1
VOLTS/DIV switch settings, up to the output-voltage upper
limit of the sine-wave generator being used.
e. Move the cable from the CH 1 OR X input connector to
the CH 2 OR Y input connector. Set the VERTICAL MODE
switch to CH 2.
f. Repeat parts b and c for all indicated CH 2 VOLTS/DIV
switch settings, up to the output-voltage upper limit of the
sine-wave generator being used.
15. Check Channel Isolation
a. Set:
VOLTS/DIV (both)
1 V
Channel 1 Input Coupling
GND
A SEC/DIV
0.1
ns
b. Set the generator to produce a 25-MHz, 5-division
display.
c. Set the VERTICAL MODE switch to CH 1.
d. CHECK— Display amplitude is 0.05 division or less.
e. Move the cable from the CH 2 OR Y input connector to
the CH 1 OR X input connector.
f. Set:
VERTICAL MODE
CH 2
Channel 1 Input Coupling
DC
Channel 2 Input Coupling
GND
5-11
Содержание 2215A
Страница 1: ......
Страница 9: ...2215A Service 4735 01 vi The 2215A Oscilloscope ...
Страница 138: ...2215A Service 4735 04 Figure 9 4 Basic block diagram ...
Страница 139: ......
Страница 145: ...CIRCUIT BCARD N TER CONNECTIONS 2 2 1 5 A 4755 20 ...
Страница 153: ...2215A S e rv ic e D UJ o C D GO CO in 10 CJ Figure 9 8 Circuit view of A1 Main board ...
Страница 157: ...1 p 2 m 3 p 4 p 5 P 6 p 7 p 8 p 9 A T B Z 2 I5 A ...
Страница 158: ...H T J t K t L t M T N T p t S 4 7JS Z2 ...
Страница 168: ...I 2 3 4 5 6 7 8 9 B H K L M N 2 2 I 5 A 473S Z ...
Страница 173: ...I 2 to 3 to 4 to 5 to 6 7 to 8 to 9 2 215A B TIMING ALTERNATE B SWEEP S ...
Страница 176: ......
Страница 181: ...I 2 3 4 5 6 7 8 9 AXIS CRT ...
Страница 184: ...2215A Service R240 DL C0MP1 C237 HF PEAK COMP R874 ASTIG R870 GEOM A1 MAIN BOARD ADJUSTMENT LOCATIONS 4735 31 ...
Страница 187: ...R652 DELAY END 4735 34 A 5 A L TE R N A T E SW EEP B O A R D A D J U S T M E N T L O C A T IO N S ...
Страница 189: ...2215A Service 4735 11 ...
Страница 193: ...3 ...
Страница 194: ...2215A SERVICE ...
Страница 195: ...4 7 5 ...
Страница 200: ......
Страница 202: ...2215A SERVICE ...
Страница 205: ...000 2775 00D ...