Operating Procedures— 2213A Operators
Phase Difference
In a similar manner to “Time Difference Between Two
Time-Related Pulses” phase comparison between two sig
nals of the same frequency can be made using the dual
trace feature of the 2213A. This method of phase difference
measurement can be used up to the frequency limit of the
vertical deflection system. To make a phase comparison,
use the following procedure:
1. Preset instrument controls and obtain a baseline
trace, then set the INT switch to CH 1.
2. Set both Input Coupling switches to the same posi
tion, depending on the type of input coupling desired.
CHANN EL 1
CHANNEL 2
MEASURE
TIM E FROM
® TO ( | )
HORIZONTAL
DIFFERENCE
C4207-19
3.
Using either probes or cables with equal time delays,
connect a known reference signal to the Channel 1 input
and the unknown signal to the Channel 2 input.
4.
Select BOTH VERTICAL MODE; then select either
ALT or CHOP, depending on the frequency of input signals.
The reference signal should precede the comparison signal
in time.
5.
If the two signals are of opposite polarity, press in
the Channel 2 INVERT button to invert the Channel 2
display.
Figure 4-8. Phase difference.
horizontal
Phase
horizontal
graticule
Difference
=
difference
x
(divisions)
calibration
(deg/div)
Example: The horizontal difference is 0.6 division with a
graticule calibration of 45° per division as shown in Fig
ure 4-8.
6.
Set both VOLTS/DIV switches and both Variable
controls so the displays are equal in amplitude.
7.
Adjust the TRIGGER LEVEL control for a stable
display.
8.
Set the SEC/DIV switch to a sweep speed which
displays about one full cycle on the waveforms.
9.
Position the displays and adjust the SEC/DIV Vari
able control so that one reference-signal cycle occupies ex
actly 8 horizontal graticule divisions at the 50% rise-time
points (see Figure 4-8). Each division of the graticule now
represents 45° of the cycle (360°
h
- 8 divisions), and Jthe
horizontal graticule calibration can be stated as 45o /per
division.
10.
Measure the horizontal difference between cor
responding points on the waveforms at a common horizon
tal graticule line (50% of rise time) and calculate the phase
difference using the following formula:
Substituting the given values into the phase difference
formula:
Phase difference = 0.6 div x 45°/div = 27°.
More accurate phase measurements can be made by us
ing the X I0 Magnifier function to increase the sweep speed
without changing the SEC/DIV Variable control setting.
EXAMPLE: If the sweep speed were increased 10 times
with the magnifier (X I0 Magnifier button out), the magni
fied horizontal graticule calibration would be 45°/division
divided by 10 (or 4.5°/division). Figure 4-9 shows the
same signals illustrated in Figure 4-8, but magnifying the
displays results in a horizontal difference of 6 divisions
between the two signals.
Substituting the given values in the phase difference
formula:
Phase difference = 6 div x 4.5°/div = 27°.
Содержание 2213A
Страница 1: ...Tektronix 2213A OSCILLOSCOPE OPERATORS INSTRUCTION MANUAL ...
Страница 5: ...2213A Operators The 2213A Oscilloscope IV ...
Страница 13: ...General Information 2213A Operators 1 8 ADD DEC 1983 Figure 1 2 Physical dimensions of the 2213A Oscilloscope ...
Страница 36: ...2213A Operators Ifektronix COMMITTED TO EXCELLENCE b I Copyright 1983 Tektronix Inc All rights reserved ...
Страница 37: ......