![Tektronix 2213A Скачать руководство пользователя страница 10](http://html.mh-extra.com/html/tektronix/2213a/2213a_instruction-manual_1078829010.webp)
General Information— 2213A Operators
Table 1-1 (cont)
Characteristics
Performance Requirements
Supplemental Information
PROBE ADJUST
Output Voltage of PROBE ADJUST
Jack
0.5 V ±5% .
Repetition Rate
1 kHz ±20%.
Z-AXIS INPUT
Sensitivity
5 V causes noticeable modulation.
Positive-going input decreases
intensity.
Useable frequency range is
d c to 10 MHz.
Maximum Safe Input Voltage
30 V (dc + peak ac) or 30 VC p-p ac at
1 kHz or less.
Input Resistance
lO k fl ±10%.
POWER SOURCE
Line Voltage Ranges
90 V to 250 V.
Line Frequency
48 Hz to 440 Hz.
Maximum Power Consumption
40 W (70 VA).
Line Fuse
1.0 A, 250 V, slow-blow.
CATHODE-RAY TUBE
Display Area
80 by 100 mm.
Standard Phosphor
P31.
Nominal Accelerating Voltage
14 kV.
REV DEC 1983