Options and Accessories
3–6
1503C MTDR User Manual
Before Testing
During Testing
Ethernet
Female type N
To 1503C
Front Panel
To 1503C
Front Panel
Female to Female
BNC to BNC
50
Ω
terminator
50
Ω
terminator
Female to Female
BNC to BNC
Female type N
Ethernet
Figure 3–3: N-Type Female T-Connector
1. Before removing the Ethernet cable terminator, make sure you have the correct
adapters and cables ready.
2. Set the 1503C front-panel controls:
CABLE
see below
IMPEDANCE
50
NOISE FILTER
1 avg
VERT SCALE
see below
DIST/DIV
appropriate setting for cable length
PULSE WIDTH
2 ns or 10 ns *
Vp
to cable specifications
POWER
ON (see below)
CAUTION.
* DO NOT use the Auto pulse width mode. The longer pulses will cause
problems on working networks.
3. Request the system administrator to notify network users of possible
disruptions.
4. Using the POSITION control, access the Ethernet Menu.
5. Scroll to 50
W
DC Termination is: Off and turn it On.
6. Return to normal operation.
7. As previously described, connect one end of a 50 jumper cable to the front-
panel CABLE connector, then connect the other end to one side of the BNC
T-connector (see Figures 4-2 and 4-3).