
Operator Tutorial
2–10
1503C MTDR User Manual
O
N
ac
50.00 ft
O
F
F
O
F
F
O
F
F
Figure 2–15: Noise Reduced.
5. Increase the NOISE FILTER setting to 128.
NOTE
. The higher the setting, the more time the instrument takes to average the
waveform.
O
N
ac
50.00 ft
O
F
F
O
F
F
O
F
F
Figure 2–16: Noise Reduced to Minimum
6. Move the
n
o
POSITION control and notice how averaging restarts at a low value
to allow easy positioning.
The 50 terminator was used here because it gives a good impedance match.
Because there are no large discontinuities, it appears to the instrument as an endless
cable. The noise seen in this demonstration is noise picked up on the cable and a tiny
amount of internal noise in the 1503C. When testing cables, the noise filter is
extremely effective in reducing noise.
Содержание 1503C
Страница 4: ......
Страница 92: ...Appendix A Specifications A 6 1503C MTDR User Manual...
Страница 104: ...Appendis B Operator Performance Checks B 12 1503C MTDR User Manual...
Страница 122: ...Glossary Glossary 6 1503C MTDR User Manual...
Страница 127: ......
Страница 128: ......