![Tektronix 067-0521-00 Скачать руководство пользователя страница 34](http://html.mh-extra.com/html/tektronix/067-0521-00/067-0521-00_instruction-manual_1078800034.webp)
Calibration— 0 6 7 -0 5 2 1 -0 0
HIGH— 600 kc ± 2 0 % (1.2 cycles/cm at 2 ftsec/
cm)
b. Set the TEST FUNCTION switch to -PULSE and
check for an inverted display.
c. Return the switch to +PULSE.
15. Check AMPLITUDE Control
a. Turn the AMPLITUDE control fully clockwise.
b. Check for approximately 6 cm of vertical deflection
(5.6 cm minimum).
c. Turn the AMPLITUDE control fully counterclockwise
and check for approximately 2 cm of deflection
(2.8 cm maximum).
16. Check Pulse Risetime
NOTE
The pulse risetime is not adjustable and is there
fore determined by the circuit parameters and the
switching characteristics of gallium arsenide diodes
D64, D65, D74 and D75.
a. Set the REPETITION RATE switch to MED.
b. Adjust the AMPLITUDE control for 4 cm of vertical
deflection.
c. Turn off the indicator oscilloscope power and re
move the 067-0521-00.
d. Insert the special plug-in extension (item 10 under
“ Equipment Required” ) in the plug-in jack.
e. Plug the 067-0521-00 into the extender.
f. Turn on the indicator oscilloscope power. (If the
indicator oscilloscope has a plug-in sensing switch
in the plug-in compartment, pull forward on the
switch plunger.)
g. Place the coupling capacitors (item 9 under “ Equip
ment Required” ) on the Input connectors of the sam
pling oscilloscope.
h. Connect the coaxial cable from pin 1 of the inter
connecting jack to the capacitor on the Channel A
Input.
i. Connect the cable from pin 3 to the capacitor on
the Channel B Input.
j. Display the Channel A signal on the sampling oscil
loscope.
k. Set the Channel A M v/Cm switch and Variable con
trol for 2.5 cm of vertical deflection.
l
. Switch to Channel B and adjust for 2.5 cm of ver
tical deflection.
m. Set the sampling oscilloscope for a differential dis
play (A + B , B inverted).
n. Center the display and readjust the Variable M v/Cm
controls slightly to produce 5 cm of vertical deflec
tion.
o. Check for a 10% to 90% risetime of 3 nsec or less
(see Fig. 6-3).
Fig. 6 -3 . Pulse Generator risetime check.
6 -4
Содержание 067-0521-00
Страница 10: ...NOTES ...
Страница 24: ...NOTES ...
Страница 46: ...CIRCUIT DIAGRAM x x r s 3 O z z l ...
Страница 47: ...J 067 0521 00 CALIBRATION FIXTURE ...
Страница 49: ......
Страница 50: ...TYPE 067 0521 00 TENT SN 2600 PARTS LIST CORRECTION CHANGE TO D80 152 0282 00 1N972B 0 4 w 30 V Mil 191 1167 ...
Страница 51: ...I ...
Страница 52: ...TYPE 0 6 7 0 5 2 1 0 0 CHANGE R5 f PARTS LIST CORRECTION 321 0152 00 37 Q 1 8 w M12 536 967 ...
Страница 53: ......