Calibration— 067-0521 -00
1. Remove signal leads from
pins 1 and 3.
2. Drill two 5 / 16-inch holes
in side of casing
3. Place
rubber
grommets
(348-0003-00; around
“ pig
tails.”
4. Insert
“ pigtails”
through
holes in casing.
5.
Solder both outer conduc
tors to short piece of bus
wire.
6 . Solder Bus W ire to pin 2.
7. Solder center conductors to
Pins 1 and 3.
♦Flexible
16-pin
plug-in
Extension ( 0 12-0038-00) may
be substituted.
Fig. 6 -1 . Construction of special extender for checking pulse risetime. Tektronix part numbers are given in parentheses.
b. Turn on the instrument power and allow 10 minutes
for warmup.
c. Set the TEST FUNCTION switch to LOW LOAD.
d. Connect the voltmeter between ground and the
+2 2 5 V front-panel jack.
e. Press the +2 2 5 V pushbutton and check for + 225
volts; then check for zero volts with the pushbutton
released.
Fig. 6-2. Left side of 0 6 7 -0 5 2 1 -0 0 showing calibration adjustments
and test points.
4. Check VARIABLE Control
a. Reset the 067-0521-00 controls to the positions given
under “ Preliminary Instructions” .
b. Set the Time Base A triggering control to free-run
the sweep.
c. Adjust the Focus and Intensity controls for a proper
display.
d. Connect a 50-ohm coaxial cable from the Calibrator
to the 067-0521-00 EXT INPUT connector.
e. Set the Calibrator for an output amplitude of 0.5
volt.
f. Turn the 067-0521-00 VARIABLE control fully clock
wise and check for approximately a 1-cm free-run
ning Calibrator signal on the crt.
g. Turn the VARIABLE control fully counterclockwise
and check that no signal is visible on the trace.
5. Check Internal Plug-In Trigger
a. Set the indicator oscilloscope Trigger Source switch
to the External position.
b. Set the Calibrator amplitude for 5 volts.
c. Adjust the 067-0521-00 VARIABLE control for 4 cm of
vertical deflection on the crt.
d. Connect the test oscilloscope probe to pin 5 of the
interconnecting plug and check for at least 65 mv
of Calibrator signal. (This signal simulates the
internal trigger from a dual-trace plug-in unit with
single-channel pickoff.)
e. Remove the probe.
6-2
Содержание 067-0521-00
Страница 10: ...NOTES ...
Страница 24: ...NOTES ...
Страница 46: ...CIRCUIT DIAGRAM x x r s 3 O z z l ...
Страница 47: ...J 067 0521 00 CALIBRATION FIXTURE ...
Страница 49: ......
Страница 50: ...TYPE 067 0521 00 TENT SN 2600 PARTS LIST CORRECTION CHANGE TO D80 152 0282 00 1N972B 0 4 w 30 V Mil 191 1167 ...
Страница 51: ...I ...
Страница 52: ...TYPE 0 6 7 0 5 2 1 0 0 CHANGE R5 f PARTS LIST CORRECTION 321 0152 00 37 Q 1 8 w M12 536 967 ...
Страница 53: ......