![Tektronix 067-0521-00 Скачать руководство пользователя страница 26](http://html.mh-extra.com/html/tektronix/067-0521-00/067-0521-00_instruction-manual_1078800026.webp)
Performance Check— 0 6 7 -0 5 2 1 -0 0
Sweep M agnifier
Horizontal Position
Intensity, Focus, Astigmatism
Scale Ilium
Trace Separation
Crt Cathode Selector
(at rear of scope)
Triggering Level
Mode
Slope
Source
Time/Cm
Time/Cm Variable
O ff
Centered
Trace focused, normal b rilli
ance
As desired
0
External Crt Cathode
Push in and rotate fully
clockwise
Auto
+
Norm
0.5 ms
Calibrated
3. Check + 2 2 5 V
a. Requirement— Positive 225 volts must appear at the
+2 2 5 V output jack on the front panel of the 067-0521-00.
b. Connect the DC voltmeter between ground and the
+2 2 5 V output jack.
c. Push the button adjacent to the + 2 2 5 V output jack.
Read the voltage on the meter and release the button.
d. Check— Positive 225 V at the output jack, which returns
to zero when the button is released.
NOTE
The v o lta g e re a d in g a t the + 2 2 5 V jack should
be the same as the test oscilloscope p o w e r supply
re a d in g .
1. Check LO W LOAD Resistances
a.
Requirement— Resistance must be within ranges indi
cated in Table 5-1.
4. Check VARIABLE
a.
Requirement— Control must be able to vary a 0.5 V
peak to peak signal amplitude from zero to a display am pli
tude of 0.6 cm or more.
b. Connect the ohmmeter between the chassis of the 067-
0521-00 and the interconnecting plug pins indicated in Table
5-1.
c. Check— That the ohmmeter reading for each pin lies
between the values listed in Table 5-1 for that pin.
TABLE 5-1
Pin N o.
M e te r Range
Resistance Reading
9
XT o k a
35 kO to 43 kO
12
X I00 k!2
infinite
101
x i ka
9.5 kO to 11.7 kO
l l 1
x i
ka
13 k a to 16 ka
15
x io o o
450 12 to 55012
2. Check HIGH LOAD Resistances
a. Requirement— Resistance must be within ranges indi
cated in Table 5-2.
b. Set the Type 1M1 TEST FUNCTION switch to HIGH
LOAD.
c. Check— That the ohmmeter reading for each pin lies
between the values listed in Table 5-2 for that pin.
TABLE 5 -2
Pin N o.
M eter Range
Resistance Reading
15
X I00 12
450 12 to 55012
12
x i ka
12.5 k!2 to 15.5 k!2
l l 1
x i ka
1.3 k!2 to 1.6 k!2
101
XI k!2
2 k!2 to 2.5 k!2
9
x i ka
2.2 k a to 2.6 ka
d.
Install the 067-0521-00 test oscilloscope and tighten the
securing rod. Turn the test oscilloscope power switch to on.
A llow at least 20 minutes warm up at 25° C, ± 5 ° , before
checking the instrument to the given accuracies.
1When measuring pins 10 and 11, connect a shorting strap across
D 80 to remove non-linear device from resistance measurements.
b. W ith all controls set as previously indicated, apply
a 0.5 V square wave from the standard amplitude calibrator
to the EXT INPUT.
c. Rotate the VARIABLE control from fully counterclock
wise to fully clockwise.
d. Check— A t the fully counterclockwise position, no
signal should be visible on the free-running trace; fully clock
wise, the signal amplitude should be 0.6 Me or more.
5. Check Plug-in Trigger
a. Requirement— Sufficient signal to provide a stable
display.
b. Set the standrad amplitude calibrator to 5 V and adjust
the VARIABLE control for a 5.5 cm display.
c. Set the test oscilloscope Time Base A Triggering Source
switch to Plug-In and adjust the Triggering Level control for
a stable square wave display (index mark on Triggering knob
should be near the center (0) Position).
d. Check— A stable display indicates that the 067-0521-00
internal trigger network is delivering a signal to pin 5 of the
connecting plug. (This signal simulates the internal trigger
from a dual-trace plug-in unit with single channel pickoff.)
6. Check VERTICAL POSITION
a. Requirement— A b ility of the VERTICAL POSITION con
trol to move the display up (clockwise) and down (counter
clockwise) over a full positioning range.
b. Set the standard amplitude calibrator to 10 V and
adjust Variable for 6 cm of display.
c. Check— Rotating the VERTICAL POSITION control fully
clockwise should move the display completely off the grati
cule in an upward direction; fully counterclockwise should
move the display completely off the graticule in a downward
direction.
5 -2
Содержание 067-0521-00
Страница 10: ...NOTES ...
Страница 24: ...NOTES ...
Страница 46: ...CIRCUIT DIAGRAM x x r s 3 O z z l ...
Страница 47: ...J 067 0521 00 CALIBRATION FIXTURE ...
Страница 49: ......
Страница 50: ...TYPE 067 0521 00 TENT SN 2600 PARTS LIST CORRECTION CHANGE TO D80 152 0282 00 1N972B 0 4 w 30 V Mil 191 1167 ...
Страница 51: ...I ...
Страница 52: ...TYPE 0 6 7 0 5 2 1 0 0 CHANGE R5 f PARTS LIST CORRECTION 321 0152 00 37 Q 1 8 w M12 536 967 ...
Страница 53: ......