Deskew and Gain Cal Fixture Instructions
Deskew Fixture Instructions
17
4.
Position the probe tips to the solder pad locations with the appropriate
polarity. See Figure 15 on page 17.
5.
Slide the probe forward until the probe tips make contact with the GAIN
CAL solder pads.
6.
Perform the following procedure to provide probe tip contact pressure with
the GAIN CAL solder pads: See Figure 9 on page 10.
a.
Pull back lightly on the probe holder assembly.
b.
While holding the probe holder assembly back, slide the probe forward
until the probe tips stick into the GAIN CAL solder pads.
c.
Now carefully release the probe holder assembly. This will maintain
contact pressure.
7.
Perform steps 1 through 6 to calibrate a second probe at the same time. See
Figure 8 on page 9.
8.
Proceed to the
Using the Deskew Fixture to Gain Cal
on page 18.
Solder pads
Figure 15: P6249, P7240 or P7260 with tips