Deskew and Gain Cal Fixture Instructions
Deskew Fixture Instructions
15
Use this procedure to Gain Cal the P6246, P6247, P6248, P6330, and P7330
probes:
1.
Insert the probe cable into the cable retainer.
2.
Insert the probe body flush to the board between the adjacent spring and
guide post. See Figure 13 on page 15.
3.
Slide the probe to allow the probe tips to contact the underside of the
fixture’s signal and ground square pins.
4.
Repeat steps 1 through 3 to calibrate a second probe at the same time, one on
each side of the fixture.
5.
Proceed to the
Using the Deskew Fixture to Gain Cal
on page 18.
Cable retainers
Spring
Figure 13: P6246, P6247, P6248, P6330 or P7330 fixture connections
P6246, P6247, P6248,
P6330, P7330
Содержание 067-0484-02