Technology Solutions
Instruction Manual
Tek-Wave 4300C
www.tek-trol.com
5
2.3
Specifications
Accuracy
±3mm
Measuring Range
90ft (30m)
Microwave Frequency
26GHz
Maximum Pressure
Max. 580psi (4Mpa)
Process Pressure
-14.50 to 580psi (0.1 to 4Mpa)
Operating Temperature
-40 to 212°F (-40 to 100°C)
Process Temperature
-40 to 482°F (-40 to 250°C)
Medium Temperature
-40 to 300°F (-40 to 150°C)
Shock Resistance
Mechanical vibration 32.80ft/s
2
(10m/s
2
), 10 to 150 Hz
Process Connection
NPT or Flange
Electrical Connection
½” NPT (Two)
Terminal Block
18 to 22 AWG Wire
Protection Class
IP67
Output Signal
4 to 20mA with HART® Optional Modbus RS485
Fault Signal
current output unchanged.
20.5mA; 22mA; 3.9mA
Integration time
0 to 50s adjustable
Power Supply
16 to 26VDC
Power Consumption
Max 22.5mA/1w
Permissive Ripple
- <100Hz Uss<IV
- (100 to 100K) Hz Uss<I0mV
Blind Area
Antenna end
Response Time
About 1 second (depending on parameter setting)
Display
LCD
Display Resolution
1mm
Approval
CE