Instruction Manual
Tek-Bar 3800XP
www.tek-trol.com
Technology Solutions
10
3.4
Installation for Wet or Dry Gas Flow
Fig 5: Installation for Wet or Dry Gas Flow
4
Electrical Connection
This section covers all electrical connection requirement. Electrical connection of the device must
be carried out by trained; qualified specialists authorized to perform such work by the installation
site.
WARNING
•
Connect all electrical cables when the power is switched off. If the device does not
have switch-off elements, then, overcurrent protection devices, lightning protection
and/or energy isolating devices must be provided by the customer.
•
The device must be grounded to a spot in accordance with regulations in order to
protect personnel against electric shocks.
NOTE
•
Improper installation has the potential to cause injury and to damage instrument.
•
Periodically inspect the power cables, transducer cables, cable glands and the
enclosure for signs of damage. Inspect transducer installation and mounting
hardware for loose connections.
Upstream/ Downstream lengths
as per: AGA / ISO or API