Operating Instructions
4-2
10 TEGAM WAY ● GENEVA, OHIO 44041 ● 440-466-6100 ● FAX 440-466-6110 ●
4.
O
PERATING
I
NSTRUCTIONS
Warm-Up Time
If the GEMINI RF Power Meter has recently been exposed to ambient temperatures beyond its
specified operating temperature of 18 to 28 °C, allow the instrument to acclimate in the
laboratory environment for at least 24 hours prior to use.
Once properly acclimated to the laboratory environment and powered on, the GEMINI RF
Power Meter requires at least 10 minutes to stabilize.
Instrument Overview
Component
Description
OLED Display
Realtime display of measurements and reference
calculations (See Section 4.4 for more information).
RF INPUT
Input connector for RF Signal under test.
RF OUTPUT
Output connector for RF Signal under test.
USB 5V
Power/Comm
Port
Connector for USB 2.0 cable. This is used for power
and communication. (See Section 4.3 for more
information)
RF OUTPUT
RF INPUT
USB 5V Power/Comm Port
OLED Display