5 - 4
5
Chapter 5 Multi-Instrument Opration
Multiple Phase Operation
Three types of multi-phase operation are available:
1. Continuous Synchronous (Cont-Sync)
2. Trigger Synchronous (Trig-Sync)
3. Trigger Sequential (Trig-Seq)
Multiple Phase Operation
Case 1 --
Continuous-Synchronous
The master unit runs continuously. The Sync Trig pulse of the master unit forces
the free-running slave unit to reset to the first address at the phase zero of the
master unit. The waveform size of the slave unit should be the same length as
the master to avoid the discontinuity.
Output
(Slave)
Sync Trig
Pulse
(Master)
Case 2 --
Trigger-Synchronous
Both units are in TRIG mode. Upon receiving the trigger signal (external or
manual), the master unit prepares to run while the SYNC TRIG pulse of the
master unit forces the slave unit to start to run synchronously at the phase zero
of the master unit.
Output
(Master)
Output
(Slave)
Sync Trig
Pulse
(Master)
Trigger Pulse
(MAN.or EXT.)
Output
(Master)
Содержание 2711A
Страница 9: ...1 1 1 Quick Start 1...
Страница 19: ...About Arbitrary Waveforms 2...
Страница 20: ...12 1 2 About Arbitrary Waveforms 2...
Страница 26: ...Front Panel Menu Operation 3...
Страница 27: ...3 1 3 Front Panel Menu Operation 3...
Страница 34: ...Waveform Creation System 4...
Страница 35: ...14 1 4 Waveform Creation System 4...
Страница 39: ...Multi Instrument Operation 5...
Страница 40: ...5 1 5 Multi Instrument Operation 5...
Страница 45: ...5 6 5...
Страница 46: ...Performance Verification 6...
Страница 47: ...6 1 6 Performance Verification 6...
Страница 51: ...7 1 7 Remote Interface Operation 7...
Страница 68: ...7 18 7...
Страница 69: ...8 1 8 Advanced Remote Operation 8...