1.5 Basic Operation
When switching on the device and a zero point adjustment was carried out it shows
shortly „nuLL Corr“.
On-/Off-Switch
Tara:
Calling of tara function, zero point adjustment
min/max:
Showing the min- resp. max-memory in sec.
display
Store/Quit:
Calling of hold function
Set/Menu:
Calling of configuration
Max Memory:
Pressing ´max´ (key 2) shows the maximum of the measured values.
Pressing it again hides it. To clear the max memory press key ´max´ for >2 seconds.
Min Memory:
Pressing ´min´ (key 5) shows the minimum of the measured values.
Pressing it again hides it. To clear the min memory press key ´min´ for >2 seconds.
Hold Function:
By pressing ´Store/Quit´ (key 6) the last measuring value will be
held in the secondary display. Pressing it again hides it.
Tare Function:
By pressing ´Tara´ (key 3) the display will be set to 0.
All measurings from then on will be displayed relatively to the set tare value.
When tara function is activated, the arrow "Tara" appears in the display.
To deactivate tare function press ´Tara´ for >2 seconds.
Please Note: Activating/deactivating tara clears the max- & min-memories.
Zero-Point Adjustment:
If there is no pressure applied to the pressure ports the
device will display 0. If there is a permanent deviation (and device is operated under
steady conditions), a permanent zero point adjustment can be carried out.
To carry out the adjustment press key 3 for 5 seconds.
(Please note: A zero-point adjustment can only be carried out if the difference
between the value on display and the value calibrated on site is less than 2%! )
To recall the manufacturer`s calibration press button 3 for approx. 7 seconds.
Note: If a zero-point adjustment was carried out, this will be signalled by the short
displaying of „NuLL Corr“ when switching on the device.