Electrical connection
Electrical and magnetic fields often generate interfering voltage in the
measuring circuit. This interference essentially emanates from high
voltage current running parallel to the measuring lines, but can also be
caused by contactors or electric motors operating in the vicinity.
Interfering voltage can also be introduced galvanically. This
particularly occurs in cases where the measuring chain is earthed at
various points that do not have the same potential.
To avoid EMC problems, please note the following:
•
Always use shielded, low-capacity measuring cables
(all tecsis cables meet these requirements).
•
The shielded measuring cable should be grounded on both sides
.
•
Do not route the measuring cable parallel to high-voltage current and
control cables.
•
Avoid leakage fields from transformers, motors and contactors.
•
The load cell, the amplifier and the processing or display unit must
not have multiple earths. Attach all equipment to the same protective
conductor.
The plug or cable connection assignments can be found on the name
plate (Fig. 1). Unless otherwise agreed, the following assignments are
used as standard.
Содержание F9304
Страница 14: ...Betriebsanleitung Operating manual F9304 14 BD_BE 9005 a www tecsis de Bauform 2 ab 25 mm Seildurchmesser...
Страница 15: ...Betriebsanleitung Operating manual F9304 www tecsis de BD_BE 9005 a 15 9 Konformit tserkl rung...
Страница 25: ...Betriebsanleitung Operating manual F9304 www tecsis de BD_BE 9005 a 25 Shape 2 from 25 mm rope diameter...
Страница 26: ...Betriebsanleitung Operating manual F9304 26 BD_BE 9005 a www tecsis de 9 Declaration of Conformity...
Страница 27: ...Betriebsanleitung Operating manual F9304 www tecsis de BD_BE 9005 a 27...