![Tecsis E3906 Скачать руководство пользователя страница 16](http://html1.mh-extra.com/html/tecsis/e3906/e3906_operating-manual_1072875016.webp)
Bedienungsanleitung/Operating manual
E3906
Presettings for sensitiv ity w ith S1 and S2
Decimal point settings w ith S3 and S4
mV/V
100%
Anzeige
S1
S2
S 3
S4
Anzeige
0,35 – 2,1
2000
open
closed
open
open
1888,8
2,0 – 3,5
2000
open
open
closed
open
188,88
0,35 –
0,55
5000
closed
closed
open
closed
18,888
0,5 – 0,9
5000
closed
open
closed
closed
18888
0,88 – 3,5
5000
open
closed
0,35 – 1,1
10000
closed
closed
1,0 – 1,82
10000
closed
open
1,75 – 3,5
10000
open
closed
0,35 – 2,2
19999
closed
closed
2,0 – 3,5
19999
closed
open
Sensor connection: Type: 15pol. SUB-D HD (High Density)
Pin 1
excitation -
Pin 11
Pin 8
exci
Pin 12
signal -
Pin 10
control
Pin 13
shield
Example 1 of calibration: w ith calibration control
Torque sensor 50 Nm, 2 mV/V.
Display at 100% = 50,00, in this case decimal point setting: S3 closed, S4 open.
Presetting for 5000 and 2 mV/V, S1 is open and S2 is closed.
Then adjust zero-potentiometer = 0 at unloaded sensor.
Press control and adjust 50,00 with amplification potentiometer.
Example 2 of calibration: w ith nominal load (100%) and w ithout calibration control
Force sensor 10N, 1mV/V.
Display at 100% = 10,000, in this case decimal point setting: S3 open, S4 closed.
Presetting for 10000 and 1mV/V, S1 is open and S2 is closed.
Then adjust with zero-potentiometer = 0 at unloaded sensor.
Apply nominal load(100%) and adjust 10,000 with amplification potentiometer.
Example 3 of calibration: w ith e.g. 123kg part load and w ithout calibration
Force sensor 200kg, 1,6mV/V.
Display at 100% = 199,99, in this case decimal point setting: S3closed, S4open (high
resolution). Presetting for 19999 and 1,6 mV/V, S1 and S2 are closed.
200,0 so decimal point setting: S3 open, S4 open (can measure more than 200kg).
Presetting for 2000 and 1,6 mV/V S1 is open and S2 is closed.
Then adjust with zero-potentiometer = 0 at unloaded sensor.
Apply 123 kg load and adj ust 123,00 (123,0) w ith amplification potentiometer.
16