Bedienungsanleitung/Operating manual
E3906
3 Configuration
Sw itch on/off:
P/ON
The device is switched on with the P/ON switch. It switches off automatically 10-15
minutes after the last activation of the P/ON switch.
Peak-lev el-detector:
P/ON
The peak-level-detector and the appropriated LED is switched on with the P/ON switch.
From this point the highest occurred measurement value is displayed. The function will
be switched off by pressing again and then be activated by immediate pressing.
Signal rev erse function:
+/-
The signal reverse function and the appropriated LED are switched on by the +/- switch.
By this function through the peak-value-detector a negative measured value is
converted into a positive measured value in order to determine the peak value.
Repeated pressing switches the function off again.
Sw itching of calibration control: CAL
100 % control can be switched on in the sensor (if available, see data sheet of the
sensor) by this key. Therewith, simple control of the system is possible. By pressing the
key, the characteristic value of the sensor will be shown at he display.
Calibration to a sensor:
Switch on device, connect sensor, signal reverse and peak-level-detector has to be
switched off.
Determine sensor sensitivity (mV/V) and 100% value (e.g. 500 Nm) and carry out the
basic settings according of the table.
Unload sensor and adjust the display to 0 with the potentiometer zero.
Switch to control and keep it pressed, or apply nominal load (100%), adjust with
potentiometer amplification 100% value (e.g. 500,0). Check zero point once more,
finished.
BD_BE 949 b