Reference Manual Rev 3.23
Software Versions 3.xx
Page 65
12.5. Error Messages
A number of error messages may be displayed to warn of operation outside of the
acceptable limits. These messages are described below. Short messages
(XXXXX) will appear as a single message on the display. Longer messages
(XXXXX) (YYYYY) will appear on the display in two parts, first the (XXXXX) part,
then the (YYYYY) part.
12.5.1.
Weighing Errors
These messages show status messages or errors that may occur during normal
weighing operation.
Error
Description
Resolution
(U - - - - -)
The weight is below the minimum
allowable weight reading.
Increase the weight or
decrease the minimum
allowable weight
reading.
(O - - - - -)
The weight is above the maximum
allowable weight reading.
Warning - overloading may damage
mechanical scale elements.
Check the condition of
load cell connections.
Check for damaged load
cell.
(ZERO)
(ERROR)
The weight reading is beyond the limit set
for Zero operation. The operation of the
<ZERO> key is limited in the setup during
installation. The indicator cannot be
Zeroed at this weight.
Increase the Zero
Range (Z.RANGE) or
use the <TARE> key
instead.
(STABLE)
(ERROR)
Scale motion has prevented a <ZERO>
or <TARE> operation from occurring on
command.
Try the operation again
once the scale is stable.