Reference Manual Rev 3.23
Software Versions 3.xx
Page 32
HI.RES (High Resolution x 10 mode)
⊗
Sets the instrument to display weight at 10 times resolution. This is intended for
test purposes in trade applications but may be used for industrial weighing. This
mode is indicated by the flashing of the units annunciator.
•
Options are: ON or OFF
•
Default: OFF
CABLE (4-Wire or 6-Wire)
⊗
Sets the load cell input to operate in 4-wire (auto sense) or 6-wire mode.
•
Options are: 4 or 6
•
Default: 6
7.4.2. OPTION (Scale Options)
Items within this Group are used to configure the operating parameters of the scale.
USE (Scale Use)
⊗
This is where the basic use of the scale is set. This setting configures the
instrument for Industrial, NSC, OIML, or NTEP operation
.
Refer to Industrial vs
NSC, OIML and NTEP Mode page 27 for more information.
•
K302 & K305: options are: INDUST (Industrial), NSC, or OIML
•
K303: options are: INDUST (Industrial), or NTEP
•
Default: INDUST
FILTER (Reading Average)
The instrument can average a number of consecutive readings when calculating
the displayed weight. This is used to dampen unwanted weight fluctuations
caused by vibrations or dynamic forces. High settings will stabilise the display
at the expense of rapid response to sudden weight changes.
•
Options are: NONE, 0.2, 0.5, 1.0, 2.0, 3.0, 4.0 (time in seconds)
•
Default: 0.5 (seconds)
MOTION (Motion Detection)
⊗
Sets how much weight variation over a defined time period is allowed before the
displayed weight is deemed to be unstable. This value is displayed as weight
change (0.5 or 1.0 graduations) per second. When set to OFF, the Motion
Detection is ignored and ZERO, TARE and PRINT actions are instantaneous.
•
Options: OFF, 0.5-1.0, 1.0-1.0, 0.5-0.5, 1.0-0.5 (graduations per
second)
•
Default: 0.5-1.0 (0.5 graduations per second)
INIT.Z (Initial-Zero on Startup)
This function can be used to automatically ZERO the indicator during power-up.
The amount of weight that can be zeroed is limited to +/– 10% of full scale.
•
Options are: ON or OFF
•
Default: OFF