![TECOM Pro-3200 Скачать руководство пользователя страница 17](http://html1.mh-extra.com/html/tecom/pro-3200/pro-3200_operation-manual_1072590017.webp)
Tecom Smart Portable Vibration Diagnosis Instrument operation manual
Page 17
13.
Installation Precautions
(1) The vibration gauge wire must be loosened. The wire close to the vibration gauge
body can not be tightened. Therefore, pay attention when handling the wire, and
leave some flexibility to allow the wire to move freely without affecting the
vibration.
(2) After vibration gauge installation is completed, you can shake the vibration gauge
by hand, the vibration gauge must be very solid, no shaking phenomenon,
otherwise the vibration value will increase, but also increases the frequency of
unpredictable. In practice, the vibrating gauge mounted with a magnet will be
secured with quick-drying glue.
(3) The point of installation must be thick steel, such as a motor body. Do not install on
fan covers or heat sink fins or thin iron plates because they are too thin and easy
to resonate, vibration values increase, and produce unexpected frequencies.
(4) Generally, the vibration value will be less than 10 mm / s. If the vibration of the
motor itself is not large and the vibration value is large, please re-check if the
installation point and the vibration gauge are installed securely.
(5) This vibration frequency response frequency range is 1-1130Hz.