Produzido por Tecnomotor Eletrônica do Brasil S/A - REPRODUÇÃO PROIBIDA - Eventuais erros ou defeitos favor comunicar [email protected]
239
TM 514 – COMMON RAIL TEST
Error
Code
209
Leakage in the drain valve CH1
Probable Cause
Solution
- Accumulation of impurities in the hydraulic
system.
- Bad contact at the drain valve connections.
- Drain valve o-ring broken.
- Contact the manufacturer.
Error
Code
210
Leakage in the drain valve CH2
Probable Cause
Solution
- Accumulation of impurities in the hydraulic
system.
- Bad contact at the drain valve connections.
- Drain valve o-ring broken.
- Contact the manufacturer.
Error
Code
211
CH1 data storage exceeded
Probable Cause
Solution
- 'Opt_sampling' remote configuration
parameter less than 99.
- Contact the manufacturer.
Error
Code
212
CH2 data storage exceeded
Probable Cause
Solution
- 'Opt_sampling' remote configuration
parameter less than 99.
-
Contact the manufacturer.
Error
Code
213
Contaminated fluid
Probable Cause
Solution
- Fluid was not changed within the specified
time.
- Fluid contamination.
- Remote configuration parameter “Opacity”
less than 25.
- Change the fluid.
Error
Code
255
Process stop requested by user.
Probable Cause
Solution
- User pressed the “RET.” key during a test
process.
- Restart the process.