22
DNVA208 Rev_5
PAR. 8
STERILIZATION TABLE
CYCLES
TIME OF
STERILIZATION
TIME OF
DRYING
CHARGE*
(Kg)
CHARGE**
(g)
WORK PRESS. (bar,
differential pressure)
C1
Not wrapped
134°C
6 minutes
17 minutes
1.25
50
2.1
÷
2.2
C2
Wrapped
134°C
11 minutes
17 minutes
1.25
50
2.1
÷
2.2
C3
Not wrapped
121°C
18 minutes
17 minutes
1.25
50
1.1÷1.2
C4 Wrapped
121°C
21 minutes
17 minutes
1.25
50
1.1÷1.2
C5
SPECIAL
134°C
6 minutes
2 minutes
1.25
50
2.1÷2.2
* The values indicate are the maximum load for every tray, for ferrous materials or solid or hollow part
** The values indicate are the maximum load for every tray, for porous parts
It is compulsory to wrap the devices aimed at an invasive use (for instance; Surgical instruments,
etc.) to ensure a higher sterility.
The data of this table are indicative; the choice of the sterilisation cycle must be based on the data
supplied by the producer of the object you want to sterilise.
PAR. 9
FILLING OF WATER TANK
TURN ON THE DEVICE PRESSING THE MAIN SWITCH.
Connect the suitable supplied hose to the loading plug and insert the other end (the one with the filter) inside
the demineralised or distilled water tank.
NOTE: The use of feedwater with contaminants at levels exceeding those given in the table above
can greatly shorten the working life of the device, with damage at the components of the autoclave
and then invalidate the manufacturer’s warranty.
Pressing the button "16A", the pump is automatically turned on. By sucking, the pump fills the tank.
The pump supplies water for 120 seconds; after reaching the maximum level, it is turned off.
In case during these 120 seconds the pump does not reach the maximum level, the operator must press the
button "16A" up to switching off the pump in automatic mode (for the maximum reached level).
The autoclave does not work in case the tank level in on MIN. as signalled by the LED.
The attempt to start the autoclave with MIN. level state is visualised with the message "ER2" on the display