IST-2136.IC01.01

 

Istruzione / User’s Manual / Manuel d’utilisation 

Pag.4/10

 

 

 

TECNOCONTROL S.r.l. Via Miglioli 47 SEGRATE ( MI ) ITALY 

Tel. +39 02 26922890 - Fax +39 02 2133734

 

 

 

È  possibile  usare 

bombole  monouso,  complete  di  valvola  d’erogazione

,  inoltre  è  necessario  usare 

il  kit  di 

prova 

Tecnocontrol mod. TC011 oppure TC014

 “TEST  ELETTRICO”  (

Codice  Test:

  F2,  F2,  F1,  F1)

 

permette  di  effettuare  un  test  funzionale  del  rilevatore.  Dopo 

aver messo l’impianto in sicurezza ed aver inserito il “

Codice Test

”, si spengono tutti i Led e si disattivano tutti i re-

lè.  Quindi  si  accenderanno  in  sequenza,  i  Led,  dal  giallo  al  3°  rosso.  All’accensione  dei  Led  corrisponde 
l’attivazione dei relativi relè (

relè "FAULT" con il Led giallo, relè "ALARM1" con il 1° Led rosso, relè "ALARM2" con il 2° Led rosso, relè 

"ALARM3" con il 3° Led rosso

). Alla fine, tutti i Led rimarranno accesi per 5 secondi, poi il rilevatore tornerà nelle condi-

zioni di funzionamento normale. È consigliabile eseguire questo Test ogni 6-12 mesi in base all’utilizzo. 

Nota:

 Non è possibile eseguire l’operazione se sono già accesi il 

1° e/o il 2° e/o il 3° Led Rossi

.

  

“VERIFICA CON GAS” (

Non serve Codice

)

serve per controllare la corretta risposta del rilevatore al gas e può es-

sere eseguita dopo l’installazione, ma va eseguita soprattutto durante le manutenzioni periodiche. 

NOTA

:  utilizzare  la  miscela  di  gas  sopra  indicata.  Normalmente  è  sufficiente  effettuare  la 

Verifica  della  Taratura

(

Bombola 

1% Volume 

CO

2

 

in azoto

) ma è possibile anche effettuare la 

Verifica dello Zero 

(

Bombola Aria Zero

). 

Infilare il TC011/TC014 sul portasensore, regolare il riduttore  della  bombola in modo che  il flussometro indichi da 
0,3 l/min (

vedi Fig. 

3

) Attendere circa 3 minuti e controllare che il rilevatore vada in allarme come decritto nel capito-

lo 

Funzionamento

Se il risultato fosse molto diverso, contattare il nostro servizio assistenza.

 Terminata la “

Verifica 

con Gas

”, chiudere la bombola, togliere il TC011. Considerare che progressivamente il rilevatore ritornerà alle con-

dizioni normali. 

NOTA

: In aggiunta a quanto sopra, se si usa anche l’uscita in mA, 

Infilare il TC011/TC014 sul portasensore,

 rego-

lare il riduttore della bombola in modo che il flussometro indichi 

0,3 l/min

 (

vedi Fig.3

). Controllare, che con i puntali 

del voltmetro sui Test-Point “

TEST mV” 

(TP1 e TP2)

, (

vedi Fig.2

) si raggiunga il valore in 

mV

 indicato in 

Tabella 4-Colonna 

6

 [

ovvero che l’uscita in mA aumenti fino al valore indicato nella 

Colonna 7

 

(± 0,56 mA) e la centrale, cui è collegato il rilevatore, indichi il 

valore in indicato in 

Colonna 8

]. Se il valore fosse  molto diverso, contattare  il nostro  servizio assistenza. Terminata  la 

Verifica

”, chiudere la bombola e allontanare il 

TC011/TC014

. L’uscita tornerà poi a 4 mA. 

Esempio

 

se si deve verificare un SE136 IC2, utilizzando prima la bombola con Aria Zero Pura l’uscita in mA dovrà 

essere circa 4,0 mA (cioè circa 40mV su “

TEST mV

”), poi utilizzando la bombola con CO

2

 a 1%vol (=10.000ppm) in 

Azoto, l’uscita in mA dovrà essere circa 12 mA 

(±0,56mA) 

(cioè 114÷126mV su “

TESTmV

”). 

 

 

EN

 

 DESCRIPTION 

The 

SE136 IC2

 series  are  gas detectors able to detect 

CO

2

 (

Carbon Dioxide

)  by  employing a 

NDIR  – Non dis-

persive Infrared Sensor

, and find their best application in centralized alarm systems for laboratories, food indus-

tries (Wineries, Breweries) or environments (Greenhouses) where monitoring the amount of carbon dioxide. 
The instruments is comprise of an enclosure in which the electronic circuit and the terminals are mounted; the sen-
sor is inside the sensor holder, in the lower part of the housing. 
The detector has three alarm levels, with different set-up as listed in 

Table 2

. The sealed relays are with tension free 

single pole contacts 

(SPS

T) and are also settable, NO or NC by the Dip-Switch. 

The instrument has also a 4÷20mA linear output (

S

) connectable to a remote Gas Central Unit. 

On the front panel, F1 and F2 key using for Test routine protected by a code and the 5 LED shows the functioning 
conditions. 

3rd Red LED "ALARM 3": 

 

3rd relay activation 

2nd Red LED "ALARM 2": 

 

2nd relay activation 

1st Red LED "ALARM 1": 

 

1st relay activation 

Green LED "ON": 

 

normal working condition 

Yellow LED "FAULT": 

 

the sensor should be faulty, disconnected or expired. 

NOTES ON THE AVAILABLE MODELS 

The 

SE136 IC2

 model is calibrated with a full scale of 

2% volume (20000ppm) of

 

CO

2

 

The 

Carbon dioxide 

(

synonym: Carbonic anhydride

), is an odourless, colourless gas, heavier than air, its density rela-

tive  to  air  is  1.5  and  this  gas  can  accumulate  at  the  bottom  causing  a  dangerous  lack  of  oxygen.  Its 

TWA

  is 

5000ppm = 0.5% volume

 (

Time-Weighted Average

) and 

STEL 15000ppm = 1.5% volume

 (

Short Term Exposure Limit

).

 

Carbon dioxide is a fundamental component of Earth's atmosphere in clean air is about 400ppm and is necessary 
to the life processes of plants (photosynthesis) and animals, but, for the earth's atmosphere, it is also a greenhouse 
gas.  The  carbon  dioxide  is  produced  by  combustion  processes  of  petroleum  and  coal  (thermal  power  plants  and 
vehicles),  as  a  by-product  of  fermentation  (wine,  beer  and  organic  substances),  or  emitted  by  natural  processes 
and by exhalations from the ground. Is used in agriculture or horticultural to increase the growth of plants, through 
increase of its concentration in greenhouses, etc. 

OPERATIONAL DESCRIPTION  

The NDIR sensor is temperature compensated, it can be used in polluted environments without a significant deg-

radation of performance and also works in environments with oxygen deficiency. 

The  operational  mode  of  the  NDIR  sensor  is  based  on  the  physical  principle  that  certain  types  of  gases  absorb 

certain  wavelengths  of  infrared  energy.  In  practice,  a  filament  generates  infrared  energy,  which  concentrated, 
passes  through  the  gas  to  be  measured,  and  arrives  at  the  detector.  The  detector  is  double,  the  first,  defined  as 
“active”, has an optical filter, selective to the gas to be measured, while the second, called "reference" has a filter 
with a different wavelength. The active detector is used to detect the gas, while the reference detector provides the 

Содержание SE136IC2

Страница 1: ...empo di risposta Response Time Temps de réponse 60s T90 105s dipendente dal valore misurato depend ing on the measured value dépend de la valeur mesurée Ripetibilità di zero Zero Repeatability Zéro répétitivité 1 FS Fondo Scala Full Scale fond d échelle Precisione Accuracy Précision 0 02 vol 200ppm 5 del valore misurato of measuring value de la valeur mesurée Tempo di Preriscaldo Warm up time Temp...

Страница 2: ...e e arriva al rivelatore Il rivelatore è doppio il primo definito attivo ha un filtro ottico per il gas da misurare mentre il secondo chiamato riferimento ha un filtro con una lunghezza d onda diffe rente Il rilevatore attivo viene usato per rilevare il gas mentre il rivelatore di riferimento fornisce il valore di zero La differenza di segnale tra i due rivelatori fornisce il valore di concentrazi...

Страница 3: ...ssima con alimentazione 12Vcc 10 è 50ohm mentre a 24Vcc 10 è 500ohm vedi Fig 4 Il morsetto d alimentazione sulla scheda principale è ad innesto e va sfilato per collegare i cavi Prestare atten zione nel reinserirlo dato che è polarizzato Invece quelli dei Relè d allarme posti sulla scheda uscite sono fissi I relè di tipo sigillato hanno un contatto libero da tensione SPST 1 Singolo Contatto Single...

Страница 4: ...la con CO2 a 1 vol 10 000ppm in Azoto l uscita in mA dovrà essere circa 12 mA 0 56mA cioè 114 126mV su TESTmV EN DESCRIPTION The SE136 IC2 series are gas detectors able to detect CO2 Carbon Dioxide by employing a NDIR Non dis persive Infrared Sensor and find their best application in centralized alarm systems for laboratories food indus tries Wineries Breweries or environments Greenhouses where mo...

Страница 5: ...e to workers OSHA Occupational Safety and Health Administration of the U S Department of Labour and COSHH Control Of Substances Hazardous to Health in Europe FAULTS the Yellow LED FAULT illuminates the S output falls down to 0mA The different faults are listed below and the FAULT normally activated relay NC contact will deactivate The FAULT relay if necessary can be used both to signal remotely an...

Страница 6: ...ith the yellow Led relay ALARM1 with the 1st red Led relay ALARM2 with the 2nd red Led relay ALARM3 with the 3rd red Led At the end all the Led will remain lighted for around 5 seconds then the detector returns at the conditions of normal operation It is advisable to perform this operation every 12 months according to the use Note this function is not working if the 1 st and or the 2 nd and or the...

Страница 7: ...utre dite de référence comporte un filtre de longueur d onde différente et servant de référence La différence entre ces deux signaux permet de mesurer la concentration de gaz L unique cause d interférence peut être donnée par la présence de vapeur d eau qui absorbe l infrarouge Préchauffage à partir du moment où la sonde est alimentée commence le préchauffage du capteur signalé par le clignotement...

Страница 8: ...contre ceux des relais d alarmes montés sur la carte de sortie sont fixes Les relais de type solide state possèdent un contact libre de tension SPST 1 Unique Contact Single Pole Single Throw qui peut être paramétré NA NO Normalement Ouvert ou NC NF Normalement Fermé selon position du Dip Switch n 4 Tableau 3 Les Dip Switch de 1 à 3 servent à déterminer les seuils d alarmes en fonction des concen t...

Страница 9: ...V TP1 et TP2 voir Fig 2 on obtient la valeur en mV indiquée au Tableau 4 Colonne 6 ou bien que la sortie en mA augmente jusqu à la valeur indiquée dans la Colonne 7 0 56mA et que la centrale à laquelle est raccordée la sonde indique la valeur contenue en Colonne 8 Si le résultat est très différent contactez notre ser vice Une fois la Vérification terminée fermer la bouteille et enlever le kit TC01...

Страница 10: ...a Gas heavier than air gaz plus lourd que l air Fig 1 Dimensioni Size Dimensions Fig 2 Schema di collegamento Wiring diagram Schéma S 12 24V mA 1 2 3 4 ON OFF SET ALARM 1 ALARM 2 ALARM 3 FAULT Scheda Uscite Relé Relays Output Board Sortie sur Realis TESTmA 12 24Vdc Alimentazione Power Supply Alimentation Centrale rivelazione gas Gas Central Unit Centrale de détection RL Uscita in 4 20 mA 4 20 mA O...

Отзывы: