13. TECHNICAL CHARACTERISTICS
13. Measuring specifications
Resolution: 0.1
Body temperature readings
Measurement range: 34,0/43,0°C (93.2/109.4°F)
Room temperature working range:
- standard range: 16,0/40,0°C (69.8/104.0°F)
- extended range: 10,0/45,0°C (50.0/113°F)
(1)
Accuracy:
from 34,0 to 35,9°C:
±0,3°C
from 93.2 to 96.7°F:
±0,5°F
from 36,0 to 39,0°C:
±0,2°C
(2)
from 96.8 to 102.2°F:
±0,4°F
(2)
from 39,1 to 43,0°C:
±0,3°C
from 102.3 to 109.4°F:
±0,5°F
(1)
When the FACE button is used in rooms where the temperature is in the range 10,0-15,9°C (50.0-
60.7°F) or 40,1-45,0°C (104.1-113.0°F), accuracy and operating range are not guaranteed, and the
temperature value is displayed alternately with the message “Lo.5» or “Hi.4” respectively.
(2)
ASTM E1965-98-(2016) laboratory accuracy requirements in the display range of 37 to 39°C (98 to
102°F) for IR thermometers is ±0,2°C (±0.4 °F), whereas for mercury in-glass and electronic thermome-
ters, the requirement per ASTM Standards E 667-86 and E 1112-86 is ±0,1°C (±0.2°F).
Other readings
Measurement range: 1,0/80,0°C (33.8/176.0°F)
Room temperature working range:
- standard range: 16,0/40,0°C (69.8/104.0°F)
- extended range: 5,0/45,0°C (41.0/113°F)
(3)
Accuracy:
from 1,0 to 33,9°C:
±0,3°C
from 33.8 to 93.1°F:
±1.8°F
from 34,0 to 35,9°C:
±0,3°C
from 93.2 to 96.7°F:
±0,5°F
from 36,0 to 39,0°C:
±0,2°C
(2)
from 96.8 to 102.2°F:
±0,4°F
(2)
from 39,1 to 43,0°C:
±0,3°C
from 102.3 to 109.4°F:
±0,5°F
from 43,1°C to 80,0°C: ±1,0°C
from 109.5 to 176.0°F
±1.8°F
(3)
When the HOME button is used in rooms where the temperature is in the range 5,0-9,9°C (41.0-
49.9°F) or 40,1-45,0°C (104.1-113.0°F), accuracy and operating range are not guaranteed, and the
temperature value is displayed alternately with the message “Lo.5» or “Hi.4” respectively.