5. APPLICATION NOTES 

 

Generation of micro particle free plasma samples and recovery of micro 
particles from platelet poor plasma:  
 
Ceveron

®

  MFU  500  is  designed  to  be  used  for  filtration  of  fresh,  clear, 

platelet poor plasma samples. 
For turbid plasma samples pre-filtration through another filter with larger 
pores is required. 
 
What is the efficiency for removal of micro particles ? 
 
After filtration of plasma spiked with 10,000 MP/

µ

l only 0.4% ± 0.004% of 

MPs were found in the filtrate. 
 
Recovery of micro particles from the filter plate 
 
MPs can be recovered from the filter plate by rinsing each well with 200µL 
of PBS or micro particle free plasma. Rinsing is performed by gentle 
pipetting the sample for 3 to 5 times and then pipetting off the sample. 
Recovery rates are in general > 85 % 
 
Sample Material 
 

Sample 

Pre-filtration /  
filter required 

Notes 

PPP 

NO 

Collection of PPP under 
standardized condition is required 

PRP 

YES 

Instead of pre-filtration prepare 
PPP by standard centrifugation 
procedure 

Whole Blood 

YES 

Prepare PPP by standard 
centrifugation procedure 

Lipaemic Plasma 

YES 

5 µm filter 

 

Lyophilized Plasma 

YES 

5 µm filter 

Some lyophilized plasma do not 
require pre-filtration 

 

6. REPLACING O-RING AND GASKET

 

 

An accessory kit containing an additional O-ring, gasket and Allen wrench is 
included with each unit of Ceveron

® 

MFU 500. O-ring and gasket should be 

replaced when worn or vacuum leakage is observed. 
For replacing the black gasket, the 12 screws have to be loosened by using 
the  Allen  wrench  included  in  the  accessory  kit.  The  O-ring  can  simply  be 
replaced manually. 
 
 

7. CLEANING AND STORAGE 

 

Basic preventive maintenance requires regularly cleaning of

 

Ceveron

®

 MFU 

500. 
Use a mild soap, standard laboratory  detergent, a dilute solution  of sodium 
hypochlorite,  or  ethyl  alcohol  to  clean  all  surfaces  of  the  manifold.  After 
cleaning,  rinse  off  any  residual  liquid  with  a  soft  cloth  or  paper  towel 
dampened in clean water, then wipe dry. 

 

Should  there  be  any  liquid  leakage  into  the  manifold  base  or  the  vacuum 
tubing,  it  can  be  removed  by  rinsing  the  base  with  distilled  water  while  the 
vacuum  pump  is  running  (the  pump  is  water  resistant).  After  rinsing  with 
water,  air  should  be  pumped  through  the  unit  at  least  for  2  runs  to  dry  out 
the system 

 

Other than occasional cleaning, the  Power Supply  is maintenance-free. For 
cleaning the case, use a clean, lint-free, antistatic and dry cleaning cloth. 

 

Store in a controlled environment at a temperature between 15°C and 30°C. 

 

 
 
 
 
 
 
 

Содержание Ceveron MFU 500

Страница 1: ......

Страница 2: ...onding receiver plates for collecting the filtrates and the retained particles It allows rapid collection of filtrates and retained particles for a large number of applications It is specifically desi...

Страница 3: ...ur country The voltage to be used in connection with Ceveron MFU 500 is preset to 12 V DC The correct settings for your country are preset by Technoclone as is the polarity of the voltage adapter You...

Страница 4: ...utton Press to release the vacuum after the pump has stopped g Case Contains vacuum pump and electronic parts h Manometer pressure gauge Gives the pressure during filtration in inch Hg or kPa i Tubing...

Страница 5: ...fferent voltage it has to be changed before using it together with Ceveron MFU 500 Before changing the output voltage remove the In Plug Power Supply from the AC socket Make sure that the In Plug Powe...

Страница 6: ...e the samples into the filter plate Never use the unit without receiver plate The filtrate would be lost and the manifold contaminated Remember the orientation of the plate e g well A1 on the top left...

Страница 7: ...e whole plate has to be sealed independent off the number of filled wells Make sure the foil is tightly put on the surface of the filter plate We recommend to use a fresh foil for every filtration Aft...

Страница 8: ...lyophilized plasma do not require pre filtration 6 REPLACING O RING AND GASKET An accessory kit containing an additional O ring gasket and Allen wrench is included with each unit of Ceveron MFU 500 O...

Страница 9: ...ut receiver plate or leaking receiver plates Rinse unit with distilled water followed by pumping air through the unit If problems continue after trying the suggested solutions contact sales technoclon...

Страница 10: ...1200 mA max UPA 2000 2000 mA max Ripple approx 100 mV Dimensions LxWxH UPA 1200 approx 105 x 60 x 43 mm UPA 2000 approx 118 x 62 x 39 mm without AC plug adapter respectively 10 ORDERING INFORMATION RE...

Страница 11: ...btained when the distributor or Technoclone GmbH has received pages 2 of the warranty card Notice The information in this document is subject to change without notice and should not be construed as a...

Отзывы: