background image

Tracer Gas Detection 

 

Tracer 

Gas 

consists 

of 

non-toxic, 

non-flammable, 

environmentally  friendly  (non-polluting)  mixture  of  5%  Hydrogen 
and 95% Nitrogen. For this reason, the Tracer Gas can be released 
into  the  environment  after  the  leak  detection  procedure  is 
completed.The  D740  detects  the  hydrogen  component  of  the 
Tracer gas and because hydrogen molecules are so small, it is an 
ideal gas for leak detection.  
 
The Tracer Gas is charged into an empty system at a pressure of 
approximately 75 psi (5bar). Run the engine with air conditioner on 
briefly to properly distribute the H2N2 Tracer Gas. As hydrogen is 
lighter  than  air,  always  probe  slightly  above  the  suspected  leak 
area.  Begin  with  the  detector  set  in  the 

“HI”  (4g/yr  setting).  Move 

the  probe  at  a  rate  of  approximately  3

”/second  and  as  close  as 

possible  to  the  suspected  area.  If  looking  for  an  apparently  larger 
le

ak,  check  first  at  the  7  g/yr  “Norm”  setting  or  the  “LO”  setting. 

Once the source of the leak is located and repaired, the Tracer Gas 
can  be  released  and  the  system  can  be  recharged  again  with 
refrigerant.  A  final  check  should  be  performed  using  an  SAE 
certified  refrigerant  gas  leak  detector  such  as  the  D440A  or  the 
D440A-UV.  

 

 

Leak Size Indicator 

 

The digital leak size indicator remains off normally but once a leak 
is  detected,  a  number  from  1-9  will  be  displayed  for  the  H2/N2 
Tracer Gas 

regardless of the sensitivity setting.  

 
The  number  will  continue  to  increase  or  decrease  depending  on 
the  amount  of  Tracer  Gasis  sensed.  The  maximum  value  will  be 
displayed once the leak source has been located. The table below 
can be used to approximate the size of leak 

 

Maximum # displayed 

Approx. Leak Size (oz/yr) 

1 -3 

 0.15 

4-6 

0.15 to 0.5 

7-9 

>0.5 

 

Содержание D740

Страница 1: ...Model D740 H2 Tracer Gas Leak Detector User Manual Detects H2 N2 5 95 Tracer Gas Design certified by ACSI to meet SAE Standard J2970...

Страница 2: ...4a in the HI setting and 7 g yr in the NORM setting while moving the probe at 3 second This complies with SAE Standard J2970 sections 7 2 and 7 3 Leak Testing with Tracer Gas is consistent with the te...

Страница 3: ...long a c lines and fittings no further away than 3 8 and moving at approximately 3 second If the detector alarms make smaller sweeps back and forth until the leak source can be pin pointed Note The D7...

Страница 4: ...ate of approximately 3 second and as close as possible to the suspected area If looking for an apparently larger leak check first at the 7 g yr Norm setting or the LO setting Once the source of the le...

Страница 5: ...in the center of the screw cap to expose the vent hole in the top of the screw cap see fig below 2 Turn on the detector and allow the unit to complete the warm up cycle 3 Place the sensor close to th...

Страница 6: ...er battery orientation Sensor Replace Filter Unscrew sensor tip as shown to replace filter Replace filter whenever it becomes visibly dirty or every 2 to 3 months depending on use Replace Sensor Remov...

Страница 7: ...itive response Product Specifications Model D740 Name Leak Detector H2 Tracer Gas Sensitivity 4 g yr R 134a equivalent Sensor Life 300 hours Response Time Instantaneous Power Supply 4 AA Alkaline batt...

Страница 8: ...with return address phone number and or email address SERVICE CENTER 2651 W 81st Street Hialeah FL33016 WARRANTY POLICY The D740H2 N2 Tracer Gas Leak Detector is warranted to be free of defects in ma...

Отзывы: