background image

4 • MN02004

Issue/Rev. 0.3 (8/02)

Section 2 – Service (continued)

3.  Disconnect output wiring and remove preamp from 

 

  junction box, Figure 6.
4.  For reassembly, reverse procedures.

 

Figure 6 

Removal and Servicing of Meter lnternals

1.  Drain product from line. Drain in direction of flow 

 

  (if possible).
2.  Make certain power has been disconnected to 

 

 meter.

3.  Remove junction box cover, disconnect output 

 

  wiring and conduit where required.
4.  Make certain meter is properly supported before 

 

  re-moving bolts and nuts.

5.  With meter removed, check straightening 

 

  section to make certain tube bundle is clean 

 

  and free from damage.
6.  If necessary to remove pick-up coil and 

 

  preamplifier, refer to procedures outlined above.
7.  Remove cotter pin and slotted nut from 

 

  downstream end of meter, Figures 7 and 8.

 

Figure 7

 

Figure 8 

8.  Remove downstream stator, Figure 9. In a 

 

  unidirectional turbine meter, the stator is located 

 

  by a Woodruff key. Care should be taken that the 

 

  Woodruff key is not lost. In a bidirectional turbine 

 

  meter, there is no Woodruff key; the two stators 

 

  are located by two deflector rings.

 

 

Figure 9 

9.  Inspect the holes in the downstream stator fins 

 

  on the thrust bearing end, Figure 10, to make 

 

  certain they are open and the inside of the 

 

  stator is clean. Inspect thrust washer for signs 

 

  of roughness, grooves, or cracks.

 

Figure 10

Содержание Smith Meter

Страница 1: ...on 2 Service Page 3 Section 3 Troubleshooting Guide Page 9 Section 4 Maintenance and Calibration Record Page 10 Section 5 Related Publications Page 11 MANUAL Smith Meter Turbine Meters Service Manual Models 4 through 24 Sentry Turbine Meters Bulletin MN02004 Issue Rev 0 3 8 02 ...

Страница 2: ...pressure fluctuations and that there is proper system back pressure 5 Check valve operation in the system 6 Check for air in the system After all other accessories have been checked and found to be operating normally the turbine meter should be checked Before beginning any service procedure stop product flow drain the line in the direction of flow if possible and relieve the pressure from the syst...

Страница 3: ...nterclockwise Figure 1 Figure 1 5 Remove pick up coil from meter by turning counterclockwise Figure 2 6 Check resistance of coil nominal dc coil resistance 1 020 Ω to determine if coil is faulty 7 For reassembly reverse procedure Figure 2 Caution Tighten pick up coil finger tight Over tightening will result in damage to coil Removal and Replacement of Preamplifier With Meter in Product Line 1 Remo...

Страница 4: ...ertain tube bundle is clean and free from damage 6 If necessary to remove pick up coil and preamplifier refer to procedures outlined above 7 Remove cotter pin and slotted nut from downstream end of meter Figures 7 and 8 Figure 7 Figure 8 8 Remove downstream stator Figure 9 In a unidirectional turbine meter the stator is located by a Woodruff key Care should be taken that the Woodruff key is not lo...

Страница 5: ...in most cases Bearing wear is minimal and bearing replacement is usually required only due to damaged surfaces 11 Should rotor bearing require replacing it is suggested that rotor assembly be returned to Measurement Solutions for rework See Return Procedures Page 2 Rotor bearing replacement can be done in the field by the following repair procedure a With proper tool and fixture Figure 12 the old ...

Страница 6: ... 2 22 in lb 6 8 5 6 18 UNC 2 8 ft lb 10 12 1 2 13 UNC 2 32 ft lb 16 3 4 16 UNF 2 120 ft lb 18 7 8 14 UNF 2 115 ft lb 20 1 12 UNF 2 170 ft lb 24 1 1 4 12 UNF 2 250 ft lb 3 Apply blueing or similar compound to downstream side of rotor hub that matches O D of stator Figure 15 Rotor is marked IN on upstream side of rotor 4 Slip platform bearing on shaft with approximately 1 4 of shaft showing Lubricat...

Страница 7: ... ring This condition will affect meter linearity and may be corrected by repositioning rotor bearing in rotor Repositioning is accomplished in ac cordance with procedures in Step 11 Page 5 This procedure can be repeated from the opposite side for a bidirectional turbine meter Figure 17 10 Place turbine meter in horizontal position Figure 18 Hold either up or downstream stator and move internal lat...

Страница 8: ...coil is tight If coil is tight and results are still not satisfactory and assuming prover is functioning properly it may be necessary to recheck rotor and platform bearing for possible damage during installation Proving the meter should be under the same conditions of flow rate pressure temperature and liquid characteristics as exist during normal opera tion of the meter For meter proving procedur...

Страница 9: ...that shield is grounded at instrument end only Decrease sensitivity on instrument front end Add suppression circuit to preamplifier Non Repeatability Damaged or missing blade Replace rotor Electrical noise interference See Underdelivery Damaged thrust washer Replace Deposits on the bearings Clean or use slofted bearings Plugged stator fin holes Clean Problems with the prover Check prover operation...

Страница 10: ...ter No Serial No Product Flow Rate K Factor Date Proved By Remarks Meter No Serial No Product Flow Rate K Factor Date Proved By Remarks Meter No Serial No Product Flow Rate K Factor Date Proved By Remarks Meter No Serial No Product Flow Rate K Factor Date Proved By Remarks Maintenance and Calibration Record ...

Страница 11: ...ference the appropriate bulletin number and title Turbine Meters Manual 4 thru 24 Sentry Series Installation Operation MN02003 Specifications 4 thru 20 Sentry Series SS02001 Flow Straightening Assemblies SS02007 Preamplifier PA 6 SS02012 Parts Lists 4 Sentry Series PO02013 6 Low Flow Sentry Series Model K2DF PO02014 6 Sentry Series PO02015 8 Sentry Series Model K2DH K2DT PO02016 10 Sentry Series M...

Страница 12: ...Changed Preamplifier from PA 4 to PA 6 throughout document The specifications contained herein are subject to change without notice and any user of said specifications should verify from the manufacturer that the specifications are currently in ef fect Otherwise the manufacturer assumes no responsibility for the use of specifications which may have been changed and are no longer in effect Contact Info...

Отзывы: