Proline Promass 300 EtherNet/IP
Operation options
27
5.2.3
Navigation area
If a function is selected in the function bar, the submenus of the function open in the
navigation area. The user can now navigate through the menu structure.
5.2.4
Working area
Depending on the selected function and the related submenus, various actions can be
performed in this area:
• Configuring parameters
• Reading measured values
• Calling up help text
• Starting an upload/download
5.3
Logging out
Before logging out, perform a data backup via the
Data management
function
(upload configuration from device) if necessary.
1. Select the
Logout
entry in the function row.
The home page with the Login box appears.
2. Close the Web browser.
3. If no longer needed:
Reset modified properties of the Internet protocol (TCP/IP) → 17.
If communication with the Web server was established via the default IP address
192.168.1.212, DIP switch No. 10 must be reset (from
ON
→
OFF
). Afterwards, the IP
address of the device is active again for network communication.
5.4
Addressing Ethernet-based fieldbuses
5.4.1
Setting the device address
The IP address of the measuring device can be configured for the network via DIP switches.
Addressing data
IP address and configuration options
1st octet
2nd octet
3rd octet
4th octet
192.
168.
1.
XXX
↓
↓
Can only be configured via software addressing
Can be configured via software addressing
and hardware addressing
IP address range
1 to 254 (4th octet)
IP address broadcast
255