Commissioning
Proline Promass 300 EtherNet/IP
20
4.4.2
Starting the Web browser
If JavaScript cannot be enabled:
enter http://192.168.1.212/basic.html in the address line of the Web browser. A
fully functional but simplified version of the operating menu structure starts in the
Web browser.
When installing a new firmware version: To enable correct data display, clear the
temporary memory (cache) of the Web browser under
Internet options
.
Prerequisite: The IP address of the measuring device is known.
1. Start the Web browser on the computer.
2. Enter the defined device address in the address line of the Web browser.
The login page appears.
Prerequisite: The IP address of the measuring device is not known.
1. Start the Web browser on the computer.
2. Read out the IP address via local operation (Diagnostics → Device information → IP
address).
3. Or set the upper DIP switch No. 2 to
ON
.
4. Restart the device.
5. Enter the default IP address 192.168.1.212 .
The login page appears.
If a login page does not appear, or if the page is incomplete → 30
Содержание Proline Promass 300 EtherNet/IP
Страница 35: ......