23
Temperature display
Select ºC or ºF.
Display Invalid Data
Entered Warnings
When set to ON, shows prompts when programming to warn the user if
a parameter entered is invalid.
Check for New Versions
When set to ON, a message box will be displayed when a new version
of Techne Workbench is available to download.
Program Defaults
Program defaults can be used to set up the preferred default settings for all newly created programs.
Individual programs can be edited as required.
Heated Lid
Use this to set the default heated lid temperature (in general
approximately 10ºC higher than the hottest block temperature in a
program).
Preheat Lid
Set to ON to pre-heat the lid to the set temperature before the thermal
cycling program begins.
Sample Cooling
Set to ON to cool the block to 4ºC while the lid pre-heats.
Pause
Set to ON to pause at the ambient temperature of the block before the
program starts, unless
Sample Cooling
has been set to ON. In this
case the unit will pause with the block held at 4ºC. The program will not
commence until the unit is manually un-paused.
Initial denaturation
Set to ON for the required hold time to denature the sample and/or
activate the enzyme before the thermal cycling program begins.