17
TC-PLUS SATELLITE SPECIFICATION
TEMPERATURE
Block temperature range
0 or 20ºC below ambient to 100ºC
Block uniformity
<±0.3ºC
Temperature accuracy
<±0.2ºC
Gradient availability
Across the full temperature range
Maximum gradient
30ºC (16ºC with the 384-well block)
Minimum gradient
1ºC
Column uniformity with a 10ºC gradient
<±0.3ºC
Pre-run sample cooling
Yes, 4ºC
Temperature set point adjustment
0.1ºC
HEATING/COOLING RATE
Maximum heating rate
5.0ºC/s
Minimum heating rate
0.1ºC/s
Maximum cooling rate
3.0ºC/s
Minimum cooling rate
0.1ºC/s
HEATED LID
Selectable heated lid temperature
35 to 115ºC or off
Pre-heat lid
Yes
Warm up time from ambient
<2 min
Over-temperature cut-out
Yes
Regulated heated lid pressure
Automatic
The heated lid is only operational if the block temperature is set above 35ºC
INTERCONNECTIVITY
Connection to TC-PLUS
Yes via USB port on rear of unit
Connection to PC control software
Yes via USB port on rear of unit
DIMENSIONS
Height (mm)
312
Width (mm)
276
Depth (mm)
375
Weight (kg)
14
POWER
Voltage
100-230V
Power
1000W