Calibration procedures for AM screen and FM screen are
carried out separately.
Edit the reference values from the list of the FMB-protocol
into the table in SpectroPlate.
Now perform the measurements on the FMB plate.
If the measurement results are out of tolerance, make a white calibration and try
again. If the values are still out of range, it is necessary to perform a screen calibration
on a Fogra Measuring Bar FMB.
Start with the white
calibration on the
clean white standard,
which is part of the
charging console.
Sample offset plate Fogra FMB
The 0 % and 100 %
patches are not
measured.
Protocol with list of
measurement values
Manual TECHKON SpectroPlate Chapter 1 - How to use SpectroPlate
Page 14