Section 6 Detailed Recording Setting
Setting of Physical Quantity Conversion Unit
The default setting of the physical quantity conversion unit on each channel is V.
A maximum of 8 characters is available for the physical quantity conversions unit when they are inputted
from main unit.
1.
Press the Down key to move the cursor to <CH Unit>.
2.
Press [ENTER] to show the physical quantity unit configurations on each channel at
present.
3.
Press the Up/Down key to move the cursor to the channel you want to set and press
[ENTER].
The physical quantity conversions unit input menu will appear.
You can modify a character at the underscore
bar. Press the Right/Left key to move
underscore bar and press Up/Down key to modify a character. Move the underscore
bar to
[
←
]
to change the item in the lower portion of the menu to
<BackSpace/ENTER>. And press [ENTER] to delete the input characters one by one
(Back Space).
4.
Make sure that the display shows <OK / ENTER> after finishing the input and press
[ENTER] to return to the physical quantity conversions unit input menu.
While continuing to hold [MENU] key, press [ENTER] key. The same settings can be
made on other channel simultaneously.
5.
Press the Up key to move the cursor to <OK>. And press [ENTER] to establish the
input and return to <SYSTEM MENU>.
Setting of the physical quantity conversion coefficient
The default physical quantity conversion coefficient (formula of AX+B) configured on each channel are as follows: coefficient A is 1 and coefficient B is 0. A
maximum of 10 digits (including symbols or decimal point) is available as input when they are inputted from main unit.
To convert recording numeral data to actual input voltage:
The AD conversion value is an integer ranging from –32768 to +32767 and
the value would be ±25000 when the input is ±100% in the input range settings.
The actual input voltage = AD conversion value of the data file × SLOPE of the header file
1.
Press the Down key to move to <CH UnitChange>.
2.
Press [ENTER] to show the physical quantity conversion coefficient configured on
each channel at present.
Page 27
Содержание es8
Страница 12: ...Note Page 4...
Страница 15: ...Section 2 Name Functions of Each Part Comparator Output Diagram Page 7...
Страница 25: ...Section 3 Before Recording Note Page 17...
Страница 36: ...Page 28...
Страница 41: ...Section 6 Detailed Recording Setting Note Page 33...
Страница 42: ...Note Page 34...
Страница 47: ...Page 39...
Страница 53: ...Section 9 Comparator Function Note Page 45...
Страница 57: ...Section 11 Saving Loading Settings Note Page 49...
Страница 59: ...Section 11 Manual Calibration Note Page 51...
Страница 61: ...Section 12 MISC MENU Note Page 53...
Страница 63: ...Section 13 Specifications Page 55...
Страница 65: ...Section 13 Specifications Page 57...
Страница 67: ...Section 13 Specifications Note Page 59...
Страница 68: ...Note...