background image

TRUBLUE USER’S MANUA

Water Level, Conductivity, and Barometric Pressure Data Loggers 

 

TE CONNECTIVITY SENSORS 

/// 

TRUBLUE USER’S MANUA

L             

 08/2016             

 

Page 4 

4. TruBlue Series Overview

 

TruBlue represents the leading edge of pressure sensing technology. Built with power conserving 
microcomputer circuitry and offering the highest level of functionality and accuracy available, the TruBlue 
series is comprised of the following data logger models: 

MODEL        

MEASURES & LOGS             

  SUBMERSIBLE?         PRESSURE REFERENCE  

255 LEVEL 

Level & temperature 

 

 

Yes 

 

Absolute 

275 BARO 

Barometric Pressure & Temperature 

No 

 

Absolute 

 

 

285 CT  

Conductivity & Temperature 

 

Yes 

 

N/A 

288 CT  

Conductivity & Temperature 

 

Yes 

 

N/A 

555 LEVEL

 

Level & Temperature 

 

 

Yes 

 

Vented, Absolute 

565 LEVEL 

Level & Temperature- High Accuracy 

Yes 

 

Vented or Absolute 

 

 

585 CTD 

Conductivity, Level & Temp 

 

Yes 

 

Vented or Absolute 

  

 

KEY FEATURES: 

•  

 

User-friendly TruWare software included, free of charge 

 

Fully sealed design 

 

5-year permanent battery (at a 15-minute logging rate); 10 year for TruBlue 255 model 

 

8MB internal memory with 56 MB available internal memory for the TruBlue 255 model 

 

Rugged 316 stainless steel or titanium with polyurethane standard or ETFE cable for  

    

harsh environments 

•  

 

2-year warranty against manufacturer defects 

 

Vented or absolute reference 

 

Calibration report included with each instrument when shipped 

 

Easy set-up and data retrieval 

 

Integrated temperature measurement up to ±0.1ºC accuracy 

 

Field upgradeable firmware 

 

Quick-disconnect cable assembly 

 

RS-485 output with SDI-12 protocol  

 

Optional SDI-12 for connectivity to 3rd party SDI-12 devices 

• 

 

Optional auxiliary battery pack for rapid sampling measurement 

 

 

 

 

 

Содержание TruBlue 255 LEVEL

Страница 1: ...CONNECTIVITY SENSORS TRUBLUE USER S MANUAL 08 2016 Page 1 TruBlue User s Manual FOR WATER LEVEL CONDUCTIVITY TEMPERATURE AND BAROMETRIC PRESSURE DATA LOGGERS 255 LEVEL 275 BARO 285 CT 555 LEVEL 585 C...

Страница 2: ...1 9 General Field Installation 11 10 Cable Protection 11 11 Conductivity 12 Types of Conductivity Measurement 12 Actual Conductivity 12 Specific Conductivity 12 Calibrating the Conductivity Sensor 12...

Страница 3: ...HNICAL SUPPORT While this manual provides extensive guidance for everything from setup through maintenance we understand that there s nothing like getting an expert on the phone to provide quick answe...

Страница 4: ...L Level Temperature High Accuracy Yes Vented or Absolute 585 CTD Conductivity Level Temp Yes Vented or Absolute KEY FEATURES User friendly TruWare software included free of charge Fully sealed design...

Страница 5: ...s the software program designed specifically to manage TruBlue data loggers and the data they collect Used together a TruBlue data logger and TruWare are ideal for monitoring groundwater and surface w...

Страница 6: ...schedules for connected data loggers Extract data from a previously initiated test View save and export test data and real time readings View data as a table or dynamic self updating graph displaying...

Страница 7: ...E 852 STAINLESS STEEL SUSPENSION WIRE PART NUMBER 852 XXXXX Is a low cost suspension solution intended for use only with absolute transducers Wire is available in configurable lengths from 0 999 feet...

Страница 8: ...le in the other near the backshell While applying slight pressure to push the two parts together slowly rotate the data logger assembly in the backshell until the keys align When that happens the hous...

Страница 9: ...provides lifetime moisture protection for vented gage water level data loggers A GORE TEX plug prevents water droplets from entering the filter while allowing air to pass freely to the internal silic...

Страница 10: ...ld It features the Hirose connector to quickly link to the vent filter connector and unique USB RS 232 conversion connector at the opposite end This same cable also provides external power to the data...

Страница 11: ...able extending up Although it may be operated in any orientation anything other than this vertical position will result in a slight zero offset in the measured data If the data logger will be operated...

Страница 12: ...CTIVITY UNITS microsiemens per centimeter S cm default millisiemens per centimeter mS cm A means of expressing what the actual conductivity of a solution would be at a standard reference temperature 2...

Страница 13: ...a Meter View window displaying a conductivity reading A With the TruBlue inserted in the calibration solution hold the mouse pointer over the conductivity reading and right click to get to the Conduc...

Страница 14: ...been chemically attacked eroded or physically altered A slow degradation of the sensor is hard to spot until it s too late Cleaning should be done with a combination of soaks and rinses in a solution...

Страница 15: ...ransients can travel distances of a mile or more 3 TRANSDUCER RESPONSE TO PRESSURE LEVEL INPUT CHANGES BECOMES SLUGGISH This is usually a sign that the sensing end of the transducer has become fouled...

Страница 16: ...aning solution Growth can be reduced by utilizing the 860 00825 Anti Fouling Guard to protect the sensor You can also coat the transducer with marine grease This may be the most effective and inexpens...

Страница 17: ...766 1500 or toll free at 1 800 745 8008 during normal working hours Eastern Standard Time Be prepared to supply the following information when requesting the RMA Data logger part number Serial number...

Страница 18: ...iagrams THE 850 00857 USB TO BACKSHELL COMMUNICATION CABLE The 850 00857 USB to Backshell Communication Cable is universal to vented gage and absolute data loggers This NON SUBMERSIBLE communications...

Страница 19: ...cludes built in conversion to RS 232C and USB as well as a DC input connector The 3 foot coiled cable can remain connected to an RS 232C modem port and communicate with a PDA handheld via USB Connect...

Страница 20: ...n Diagrams Continued THE 850 00856 USB TO HIROSE COMMUNICATION CABLE The 850 00856 USB to Hirose Communication Cable is the newest and most streamlined cable offering This cable provides connection fr...

Страница 21: ...data is stored in the on board memory Polled data is sent to a network capable auxiliary logger or telemetry system that capture the live data and transmits from a remote location These devices often...

Страница 22: ...Continued THE 852 XXXXX STAINLESS STEEL SUSPENSION WIRE The 852 suspension wire assembly is intended for use with absolute transducers to suspend the unit This assembly includes a stainless steel or t...

Страница 23: ...ny Non Vented data logger as a long term deployment solution This Direct Hirose to Backshell design allows for the user to leave a non vented data logger deployed in the field while still maintaining...

Страница 24: ...droplets from entering the filter while allowing air to pass freely to the internal silica gel desiccant The filter housed in corrosion resistant grade 316 stainless steel provides surge protection a...

Страница 25: ...dix A TruBlue Connection Diagrams Continued THE 860 00825 ANTIFOULING SCREEN Extend TruBlue conductivity data logger deployments in high bio fouling environments with a TruBlue Anti fouling Screen mad...

Страница 26: ...sing out of the sale resale use or misuse of the product Users of TE Connectivity products should make their own evaluation to determine the suitability of each such product for the specific applicati...

Отзывы: