Two Possible Logical Ampli
fi
er Con
fi
gurations for a PZ5-64-2 (with two digital inputs)
Logical ampli
fi
er con
fi
gurations can be de
fi
ned using the front panel interface (see
), in Synapse, or in RPvdsEx using the PZ5_Control macro. The PZ5-0-32 model
can have a maximum of two logical ampli
fi
ers con
fi
gured. All other PZ5s can have a maximum
of four logical ampli
fi
ers.
Analog Recording Reference Modes
The PZ5 supports four referencing modes for each analog input logical ampli
fi
er: Local,
Shared, None and Differential. See
Reference
Mode
Description
Local
Each bank of channels in a logical ampli
fi
er uses its own reference input (pin 5) as the reference for
that bank
Shared
The reference (pin 5) of the
fi
rst bank of the logical ampli
fi
er acts as a reference for all banks in the
logical ampli
fi
er
None
The references for all banks of a logical ampli
fi
er are tied to the Ground (pins 13, 15, and 16).
Differential
The inputs in each bank of the logical ampli
fi
er are paired; odd channels serve as recording (+)
channels and each even channel is used as an individual reference (-) channel for the preceding odd
channel. No connections should be made to pin 5.
PZ5 NeuroDigitizer | 9