3  INSTALLATION 

3.1  INITIAL INSPECTION 

The shipping container should contain the following items: power supply, HV output ca-
ble,  test  data  sheets  and  operator's  manual.  Examine  the  items  immediately  for  dam-
age. Locate the serial number labels on the power supply and verify the model number, 
the  input  voltage  rating  and  the  output  voltage  rating  and  polarity.  In  the  event  of  any 
damage promptly notify the transportation company and the TDK-Lambda Americas Inc. 
customer service manager. 

3.2  MOUNTING AND COOLING REQUIREMENTS 

The  power  supply  can  be  mounted  in  a  standard  19"  EIA  enclosure  or  equivalent.  
Chassis support brackets or rails must be added to the bottom of the power supply for 
proper weight distribution. The power supply can also operate on a bench or table top.  
In all cases adequate clearances must be provided for cable bends. Generally, at least 
4" (161.6mm) of clearance should be allowed at the rear of the power supply. 

When operating in an enclosed system, care must be taken to ensure the ambient air to 
the  power  supply  does  not  exceed  the  maximum  ambient  operating  temperature  of 
45°C, this may require addition of a system heat exchanger. 

The power supply requires a cooling water supply with a minimum 2GPM (7.6 l/m) flow 
and  a  maximum  inlet  temp  of  35°C.  The  water  connection  is  at  the  rear  of the supply.  
The connection is SWAGELOK 12mm. The water flow direction must be as marked on 
the rear panel. All water connections are at ground potential. The power supply can be 
operated in any orientation/position. 

3.3  GROUNDING AND INPUT AC POWER 

Proper grounding of the input AC power is required to reduce the risk of electric shock.  
The metal chassis of the power supply is grounded through the green ear thing wire at 
the  input  AC  power  terminal  block.  Use  extreme  caution  when  connecting  AC  input 
power  and  never  apply  the  incorrect  input  power.  Connect  the  three  lines  of  the  input 
power to the L1, L2, L3 terminals and the earth ground to the terminal marked with the 
protective earth symbol. No neutral connection is required for the 200V and 400V con-
figurations. Verify the input voltage configuration (208 or 400VAC) is on the nameplate 
on the rear panel before applying the correct voltage. Applying incorrect voltage will im-
mediately cause major damage and void the warranty. The input voltage change should 
only  be performed by qualified technical personnel equipped with proper parts and de-
tailed  instructions  from  the  factory.  Refer  to  Section  5  "Applications",  to  calculate  line 
currents for various operating conditions such as reduced power or charging very large 
capacitor banks. 

Содержание LC1202

Страница 1: ...94001 Rev J INSTRUCTION MANUAL FOR LC1202 HIGH VOLTAGE POWER SUPPLY 83494001 Revision J TDK Lambda Americas Inc 405 ESSEX ROAD NEPTUNE NJ 07753 TEL 732 795 4100 FAX 732 922 9334 WEB www us tdk lambda...

Страница 2: ...as defined by TDK Lambda Americas Inc for the products described by this manual is stated as follows The Power Supplies described by this manual are intended for use in a protected industrial environm...

Страница 3: ...AL VOLTAGES PRESENT All power supplies contain hazardous voltage and energy The power supply must only be operated by qualified personnel who have read this operator s manual and are familiar with the...

Страница 4: ...ive 93 68 EEC European Community Conformity Assessment Product Mark IEC 348 Attention consult Accompanying documents IEC 60417 1 5036 Dangerous voltage IEC 60417 1 5019 Protective earth e g power line...

Страница 5: ...ipment to the factory the following steps shall be taken 1 Notify TDK Lambda Americas Inc at 732 795 4100 or follow the instructions at US TDK Lambda com HP service htm Give a full description of the...

Страница 6: ...4 POLARITY 2 2 5 HV INSULATING MEDIUM 2 2 6 INPUT CONNECTIONS 2 2 7 INPUT REQUIREMENTS 2 2 8 POWER FACTOR 2 2 9 EFFICIENCY 2 2 10 STORED ENERGY 2 2 11 STABILITY 3 2 12 PULSE REPEATABILITY 3 2 13 PHYSI...

Страница 7: ...NE 8 3 6 2 OUTPUT GROUND CONNECTION 8 4 OPERATION 10 4 1 L S OEM MODELS 10 4 2 REAR PANEL 11 4 3 FRONT PANEL CONTROL L MODEL ONLY 12 4 3 1 POWER SWITCH 12 4 3 2 CONTROL SWITCH 12 4 3 3 HV ON SWITCH 12...

Страница 8: ...2 7 2 1 CHANGE SELECTOR BOARD 22 7 2 2 CHANGE FUSE 22 7 2 3 SELECTOR SWITCHES ON AUX POWER BOARD 23 7 2 4 CORRECTING THE INPUT LABEL 23 List of Figures Description Page Figure 1 LC1202 Block Diagram 1...

Страница 9: ...t regulation of the output voltage and automatically compensates for line load and temp variations The latest development resonant inverter topology which improves pulse to pulse re peatability by red...

Страница 10: ...1 35A 30KV 0 9A 2 3 1 LINEARITY To within 1 of full scale 2 3 2 ACCURACY To within 1 of rated output 2 4 POLARITY Available as fixed positive or negative 2 5 HV INSULATING MEDIUM Silicone oil 2 6 INPU...

Страница 11: ...7 COOLING WATER 2GPM 7 6 liters min max supply temp 35 C All water paths are at ground potential and are copper or brass An analog temp signal 0 10VDC is provided for a heat sink temp of 10 60 3 C at...

Страница 12: ...500Hz 0 015g2 Hz Duration 1 hour 2 24 SHOCK 25Gpk half sinewave 11ms 2 25 FRONT PANEL DISPLAY Voltage and current 3 character indicators 10 segment voltage and current trend graphs HV ON push button...

Страница 13: ...83494001 Rev H Page 5 of 27 Figure 2 Mechanical Details...

Страница 14: ...ling water supply with a minimum 2GPM 7 6 l m flow and a maximum inlet temp of 35 C The water connection is at the rear of the supply The connection is SWAGELOK 12mm The water flow direction must be a...

Страница 15: ...sert the connector end of the HV cable and tighten the locking nut only hand tight When operating above 20kV or 200 Hz rep rate it is recommended that a silicone grease such as Dow Corning DC 4 be use...

Страница 16: ...Some peak current will flow out of the power supply during discharge and return through the HV return and system chassis This current comes from voltage reversal in under damped systems and from norm...

Страница 17: ...Ground Connection This product is NOT protected against external load voltage reversal events Load voltage reversal may result in damage to the unit which is not covered by the standard product warra...

Страница 18: ...ed which is controlled by the front panel power switch The model LC1202L can be operated as a master unit in parallel with several model LC1202S slave units for increased output power Refer to Section...

Страница 19: ...ONTROL KEY SWITCH on the front panel is in the REMOTE position The INTERLOCK terminal block internal 24 DC can be con nected to system interlocks to disable the power supply when opened Jumper the two...

Страница 20: ...es the power supply regardless of other com mands 4 3 3 HV ON SWITCH Turns on the high voltage output to the level set by the HV adjust knob Also used with HV OFF to reset latched fault conditions ove...

Страница 21: ...k charging voltage Impedance 1k If the 5V option is installed the voltage level is 0 5V 10 HV ON LED Open collector through 100 Low impedance when HV output is enabled See Note 1 12 GROUND Control cir...

Страница 22: ...that are common to the remote control interface indicator signals A pullup resistor is not required to sense the remote interface voltage on these signals The user should expect the Active Logic leve...

Страница 23: ...83494001 Rev H Page 15 of 27 Figure 10 Suggested Interface...

Страница 24: ...1202L Front Panel Control 1 Turn HV ADJUST knob fully counterclockwise 2 Turn POWER switch to ON 3 Turn CONTROL key switch to LOCAL 4 Push HV ON switch 5 Verify the HV output is at approx zero volts 6...

Страница 25: ...00 J sec VOLTAGE TIME Tc Tp 5 2 VOLTAGE REVERSAL SEE ONLINE APP NOTE 517 FOR MORE DETAILS When the capacitor or PFN is discharged a high peak current may flow out of the power supply as a result of vo...

Страница 26: ...e cases it is necessary to wrap the cables several times through large ferrite cores at the rear panel of each unit 5 4 MEASURING HIGH VOLTAGES A sample of the output voltage is available in the REMOT...

Страница 27: ...AS A DC POWER SUPPLY The LC1202 can provide a regulated voltage to a DC load such as an electron beam The output current will be preset at the factory in order to limit the average output power to 12...

Страница 28: ...assembly hermetically sealed Opening the assembly may compromise performance 6 4 TROUBLESHOOTING First check for obvious trouble such as input power output connections control connec tions and signal...

Страница 29: ...83494001 Rev H Page 21 of 27 Check the V PROGRAM input at U1 12 Check the clock at Drain Q4 Check the transistor gate drives at Q4 Q5 Q8 Q10...

Страница 30: ...upply is disconnected from the line and load and all caps must be discharged Take the cover off On the inverter board Figure 11 discharge the bus capacitor by shorting SWGC1 SWGC2 and SWGC3 SWGC4 with...

Страница 31: ...There are 3 selector slide switches on the Aux Board which have to be correctly selected For 208VAC line operation the position of the slide switches should be so that 110VAC is visible For 400V line...

Страница 32: ...83494001 Rev H Page 24 of 27 Figure 11 Power Supply Top View...

Страница 33: ...83494001 Rev H Page 25 of 27 Figure 12 Power Supply Bottom View...

Страница 34: ...83494001 Rev H Page 26 of 27 Figure 13 Aux Board...

Страница 35: ...83494001 Rev H Page 27 of 27 Figure 14 Fuse Assembly...

Страница 36: ......

Страница 37: ......

Страница 38: ......

Страница 39: ......

Страница 40: ......

Страница 41: ......

Страница 42: ......

Страница 43: ......

Страница 44: ......

Страница 45: ......

Страница 46: ...GS 8 16 2017 BHuhn 8 16 17 A Hague 8 16 17...

Страница 47: ......

Страница 48: ......

Страница 49: ......

Страница 50: ......

Отзывы: