TCS TrainSpeed 4 Скачать руководство пользователя страница 13

 

13 

If the lighting appears to be OK but a sensor is still ‘in the 
dark’, try swapping cables with a working set of sensors 
(power down the unit before swapping cables).  If the other set 
of sensors work, then the problem lies with the sensor cable, 
not the TrainSpeed 4 display unit.  Dismount the sensor pair 
and inspect the connections between the cable and the optic, 
and the cable and the connector.  Also look to see if the cable 
itself has been cut along its length.  If you have a damaged 
sensor assembly or the swapping of sensors did not work, 
contact TCS Inc.  for technical assistance.   
 
Another problem that can occur is the unit powers up correctly 
but a train does not trigger the sensor properly.  This can, on 
rare occasions, be caused by too much light (especially on the 
larger scales).  It can also be caused by a short circuit in the 
sensor cable assembly.  To check to see if there is too much 
light, put the unit in test mode for the sensor in question as 
outlined above, and cover the sensor completely and tightly 
with a dark opaque material (usually a hand pressed down 
firmly will do).  If you see a change in the display from ‘L’ to 
‘H” then the sensor is working (if there is no change see the 
next paragraph).  Now remove the cover and roll a train car 
across the sensor.  If you do not see the display change or it 
does not stay ‘H’ while the car is over it then there is too much 
light.  There are several ways to address this.  You can recess 
the sensor further down into the mounting hole (even below 
the roadbed surface) to try to reduce the amount of light 
hitting the sensor.  If this doesn’t work you can use some 
black paint to reduce the area of the sensor’s lens (ballast can 
be piled up around the sensor to do this too).  Sometimes the 
sensor can pick up some excess light from floor reflections, so 
try covering the underside of the mounting hole with black 
tape.  If these measures do not fix the problem contact TCS 
Inc. for technical assistance. 
 
If the above test failed, then the sensor assembly is probably 
bad.  First, try a different (previously working) sensor cable on 

Содержание TrainSpeed 4

Страница 1: ...TrainSpeed 4 Four separate train speed indicators Model Railroad Speedometer Operating Manual TCS Inc 215 453 9145 Main Number 215 257 0735 Tech Number www tcsdcc com SKU 1548...

Страница 2: ...ng with four 4 screw clearance holes If you are using a clear plexiglass control panel then you can mount the unit behind a clear window instead of cutting a hole in the panel The layout for these hol...

Страница 3: ...99 11 49 5 75 cm OO 76 1315 78 52 62 26 31 13 16 6 58 cm O 48 2083 33 83 34 41 67 20 83 10 42 cm S 64 1562 50 62 5 31 25 15 63 7 81 cm G 24 4166 67 166 66 83 33 41 67 20 83 cm TT 120 833 33 33 34 16 6...

Страница 4: ...chosen spacing watch out for switch machines and the like under the layout when choosing the location of the sensors Make sure that the edges of the holes are clean especially on the bottom as this w...

Страница 5: ...old the cables in place as they can cut the insulation and cause shorts If you must anchor the cables to the benchwork staple a cable tie to the benchwork and then wrap the cable tie around the sensor...

Страница 6: ...wall transformer specifications The unit needs at least 6V to operate correctly and the supply voltage should not exceed 18V AC or DC The unit requires less then 100ma current to function The wires f...

Страница 7: ...7 6v 18v AC or DC Input polarity does not matter Figure 1 Sensor 1 Pin 1 Sensor 2 Pin 1 Sensor 3 Pin 1 Sensor 4 Pin 1 Power Connector J5 Test jumper Option Jumpers...

Страница 8: ...talled then there is a problem with that sensor pair For example if sensor pair 2 was not connected or connected incorrectly the display would look like this 43 1 At power up the unit will scan the fo...

Страница 9: ...t digit of the display with the rest of the display blank Once the second sensor is tripped you will see the scale speed displayed on the unit 2 30 This shows that sensor 2 was activated and the train...

Страница 10: ...ut two minutes on the 1 100 jumper setting for this message to be displayed The unit will display normal speed for the next train whenever any of these errors occur If the message Er 1 is displayed it...

Страница 11: ...e you will need to power down and power up the TrainSpeed board to initialize it correctly One common occurrence with the connector used is that only one row of pins is covered by the connector If you...

Страница 12: ...y H the display would look like this L3H When a sensor is covered in the dark the test display for that sensor should read H When the sensor is lit you should see an L in the test display For the unit...

Страница 13: ...ensor in question as outlined above and cover the sensor completely and tightly with a dark opaque material usually a hand pressed down firmly will do If you see a change in the display from L to H th...

Страница 14: ...P x followed by Sn E then one of the sensor cables has been damaged Plug them in one at a time until you isolate the problem cable Inspect this cable for damage and contact TCS Inc for a replacement i...

Страница 15: ...he connector Wall transformers Radio Shack AC adaptors NTE 57 9D 600 4 or NTE 57 9D 600 2 will work with this unit Cut the circular connector off that comes with the adapter strip and tin the wires an...

Страница 16: ...16 TCS Inc 215 453 9145 Main Number 215 257 0735 Tech Number www tcsdcc com...

Отзывы: